| Literature DB >> 25860665 |
Lisa N Rascoe1, Courtney Price1, Sun Hee Shin2, Isabel McAuliffe1, Jeffrey W Priest3, Sukwan Handali1.
Abstract
Strongyloides stercoralis is a widely distributed parasite that infects 30 to 100 million people worldwide. In the United States strongyloidiasis is recognized as an important infection in immigrants and refugees. Public health and commercial reference laboratories need a simple and reliable method for diagnosis of strongyloidiasis to identify and treat cases and to prevent transmission. The recognized laboratory test of choice for diagnosis of strongyloidiasis is detection of disease specific antibodies, most commonly using a crude parasite extract for detection of IgG antibodies. Recently, a luciferase tagged recombinant protein of S. stercoralis, Ss-NIE-1, has been used in a luciferase immunoprecipitation system (LIPS) to detect IgG and IgG4 specific antibodies. To promote wider adoption of immunoassays for strongyloidiasis, we used the Ss-NIE-1 recombinant antigen without the luciferase tag and developed ELISA and fluorescent bead (Luminex) assays to detect S. stercoralis specific IgG4. We evaluated the assays using well-characterized sera from persons with or without presumed strongyloidiasis. The sensitivity and specificity of Ss-NIE-1 IgG4 ELISA were 95% and 93%, respectively. For the IgG4 Luminex assay, the sensitivity and specificity were 93% and 95%, respectively. Specific IgG4 antibody decreased after treatment in a manner that was similar to the decrease of specific IgG measured in the crude IgG ELISA. The sensitivities of the Ss-NIE-1 IgG4 ELISA and Luminex assays were comparable to the crude IgG ELISA but with improved specificities. However, the Ss-NIE-1 based assays are not dependent on native parasite materials and can be performed using widely available laboratory equipment. In conclusion, these newly developed Ss-NIE-1 based immunoassays can be readily adopted by public health and commercial reference laboratories for routine screening and clinical diagnosis of S. stercoralis infection in refugees and immigrants in the United States.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25860665 PMCID: PMC4393093 DOI: 10.1371/journal.pntd.0003694
Source DB: PubMed Journal: PLoS Negl Trop Dis ISSN: 1935-2727
Performance of SS-NIE-1 ELISA and Luminex.
| Assay Characteristics | SS-NIE-1 ELISA | 95% CI | Ss-NIE-1 Luminex | 95% CI |
|---|---|---|---|---|
| Cut-off point | 0.8 ng IgG4/ mL | 8 MFI units | ||
| J-index | 0.88 | 0.88 | ||
| Sensitivity: | 245/258 (95) | 92–97 | 162/175 (93) | 88–96 |
| Specificity: | 303/325 (93) | 90–96 | 349/366 (95) | 93–97 |
Note: * CI = confidence interval
Cross-reactivity of SS-NIE-1 ELISA and Luminex.
|
|
| |||||
|---|---|---|---|---|---|---|
| Cross reactivity (%) | No. of positives | No. of sera tested | Conditions represented by sera | No. of sera tested | No. of positives | Cross reactivity (%) |
|
|
|
|
|
|
|
|
|
| NT | NT |
| 2 | 1 | 50 |
|
| 1 | 14 |
| 8 | 1 | 13 |
|
| 0 | 1 |
| 1 | 0 | 0 |
|
| 1 | 2 |
| 2 | 0 | 0 |
|
| 0 | 2 |
| 4 | 0 | 0 |
|
| NT | NT |
| 1 | 0 | 0 |
|
| 1 | 2 |
| 5 | 0 | 0 |
|
| 0 | 2 |
| 1 | 0 | 0 |
|
| 6 | 63 |
| 63 | 6 | 10 |
|
| 2 | 4 |
| NT | NT | NT |
|
| 2 | 6 |
| 14 | 2 | 14 |
|
| NT | NT |
| 3 | 1 | 33 |
|
| NT | NT |
| 6 | 0 | 0 |
|
| 0 | 4 |
| 5 | 0 | 0 |
|
| NT | NT |
| 4 | 0 | 0 |
|
| 4 | 33 |
| 28 | 1 | 4 |
|
| 0 | 4 |
| 3 | 0 | 0 |
|
| 1 | 4 |
| 4 | 1 | 25 |
|
| 2 | 2 |
| 4 | 1 | 25 |
|
| NT | NT |
| 1 | 0 | 0 |
NT = Not tested
Change in antibody response following treatment.
| At baseline | < 3 MONTHS | 3–6 MONTHS | 7–9 MONTHS | 10–18 MONTHS | > 18 MONTHS | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ELISA | Luminex | ELISA | Luminex | ELISA | Luminex | ELISA | Luminex | ELISA | Luminex | ELISA | Luminex | |
|
| 48 | 25 | 2 | 1 | 26 | 11 | 21 | 12 | 22 | 12 | 9 | 7 |
|
| 0–255 | 11–20321 | 0.5–9.7 | 480 | 0–89 | 3–9483 | 0.5–77 | 8–7447 | 0–33.3 | 15–6440 | 0–16.5 | 20–3324 |
|
| 28.8 | 3247 | 5.1 | 480 | 14.4 | 1663 | 16.3 | 1622 | 7.5 | 1293.3 | 3.7 | 1118 |
|
| 13.8 | 596 | 5.1 | 480 | 5.9 | 448 | 4.6 | 241 | 3.5 | 168 | 1.8 | 81 |
|
| 50 | 0 | 70 | 82 | 68 | 73 | 88 | 83 | 71 | 71 | ||
|
| 0.3–2.3 | 0.7–0.7 | 0.1–1.4 | 0.2–0.9 | 0.1–2.2 | 0.1–0.8 | 0–3.3 | 0.1–1.4 | 0–2.4 | 0.1–2.3 | ||
Note:
† For SS-NIE-1 ELISA, units are expressed in ng/mL and for SS-NIE-1 Luminex, the unit is a ratio of mean fluorescent intensity minus the background of the test/control.
*Follow-up serologic result divided by the initial result