Peritoneal dialysis in an extremely low-birth-weight infant with acute kidney injury.

Lyndsay A Harshman1, Melissa Muff-Luett1, Mary L Neuberger1, John M Dagle2, Joel Shilyansky3, Carla M Nester1, Patrick D Brophy1, Jennifer G Jetton1.   

Abstract

Critically ill neonates are at high risk for acute kidney injury (AKI). Renal supportive therapy (RST) can be an important tool for supporting critically ill neonates with AKI, particularly in cases of oliguria and fluid overload. There are few reports of RST for management of oligo-anuric AKI in the extremely low-birth-weight infant weighing <1000 g. We report successful provision of peritoneal dialysis (PD) to an 830-g neonate with oligo-anuric AKI through adaptation of a standard pediatric acute PD catheter.

Entities:  

Keywords:  acute kidney injury; extremely low-birth-weight infant; peritoneal dialysis; renal supportive therapy

Year:  2014        PMID: 25859376      PMCID: PMC4389134          DOI: 10.1093/ckj/sfu095

Source DB:  PubMed          Journal:  Clin Kidney J        ISSN: 2048-8505


Background

Neonates, especially very low-birth-weight infants [1], are at high risk for acute kidney injury (AKI), and AKI is an independent predictor of morbidity and mortality in this patient population [1-5]. When used in neonates, renal supportive therapy (RST) is often considered a last resort when all other medical management has failed. The decision to offer RST to the smallest of patients is complicated by a number of challenges including limited availability of appropriately sized dialysis catheters, lack of machines designed for neonates, technical difficulties associated with access placement in small infants, as well as individual physician practices, biases and beliefs. Limited available literature documenting the use of peritoneal dialysis (PD) in extremely low-birth-weight (ELBW) infants describes uniformly high rates of technical complications (25–60%) and mortality (>50%) [6-8]. For these reasons, practitioners may defer RST in favor of diuretic therapy and nutritional restriction in ELBW neonates or offer no dialytic support at all. We report the successful provision of PD to an 830-g premature infant with oligo-anuric AKI and deteriorating respiratory status using a commercially available temporary PD catheter.

Case report

The patient was a male infant born at 28 and 3/7 weeks gestation weighing 830 g. He was Twin B of a monochorionic twin gestation, and the donor in a twin–twin transfusion syndrome. The pregnancy was complicated by oligohydramnios, worsening cardiac function and hydrops in Twin A necessitating preterm delivery via cesarean section. Twin B's Apgar scores were 8 and 8, at 1 and 5 min, respectively; however, he developed respiratory distress and required intubation. He was placed on high-frequency oscillatory ventilation. The infant was anuric on Day of life (DOL) 1. He received furosemide and aminophylline to stimulate urine output without effect. Metolazone was added, but he remained anuric. A renal ultrasound demonstrated echogenic kidneys with normal Doppler flow and an empty bladder. By DOL 3, the infant's weight had risen to 1050 g, a 27% increase from birth weight. Laboratory studies were significant for serum sodium 128 mEq/L (reference range, 135–145 mEq/L), serum creatinine 2.3 mg/dL (reference range, 0.2–0.9 mg/dL) and blood urea nitrogen (BUN) 69 mg/dL (reference range, 10–20 mg/dL). The infant was transferred to the University of Iowa Children's Hospital Neonatal Intensive Care Unit for respiratory and renal management. Diuretic therapy was discontinued upon arrival given lack of effect. Total fluids were restricted to a volume of 75 mL/kg/day for the first 24 h post-transfer with further restriction to 60 mL/kg/day and 1 g/kg/day of protein on DOL 4–5 due to worsening fluid overload and significant decline in respiratory status. The infant's renal parameters indicated severe anuric renal failure, with a BUN and creatinine of 87 and 2.7 mg/dL, respectively. On DOL 5, the infant's weight was 1096 g, indicating a 32% fluid weight gain from birth. Given the dire clinical picture and the likelihood of death without intervention, the neonatology, pediatric surgery and pediatric nephrology teams reviewed the risks and benefits of RST with the patient's family and devised a PD plan for the infant.

Procedure

The primary technical challenge for initiating PD in this ELBW infant was that standard cuffed PD catheters, with lengths from 31 to 39 cm, are too large for an ELBW infant's small abdominal domain. Acute PD catheters with 8 cm catheter lengths represent the smallest commercially available PD catheter. However, these are likely still too large for an 800-g infant when placed in the standard fashion via an infra-umbilical stab incision with the catheter directed into the pelvis. To use a true PD catheter rather than improvised equipment (e.g. large-bore intravenous cannula) [9] that would be prone to complications, we thus considered a novel placement approach using an acute PD catheter inserted into the left upper quadrant. An 8.5-French, 8-cm commercial temporary PD catheter (Pediatric Peritoneal Dialysis Set—Cook Medical) was placed at the bedside by a pediatric surgeon. A small stab incision was made in the left upper quadrant. A dilator was then placed through the stab incision into the abdomen. Using the Seldinger technique, the PD catheter was exchanged for the dilator and directed into the pelvis such that the catheter remained straight. A purse-string suture was used to secure a watertight closure of the posterior rectus sheath. The subcutaneous tissues and skin were closed with running suture to prevent leakage of peritoneal fluid and permit initiation of PD immediately. The exit site was closed around the catheter and dressed. Catheter placement was confirmed radiographically (see Figure 1a and b) and with successful instillation and aspiration of saline from the catheter. See Figure 2 for photograph of catheter location.
Fig. 1.

(a) AP X-ray demonstrating PD catheter placement coursing from the left upper quadrant into the left paracolic gutter. (b) Lateral X-ray demonstrating PD catheter placement coursing from the left upper quadrant into the left paracolic gutter.

Fig. 2.

Photograph of PD catheter placement.

(a) AP X-ray demonstrating PD catheter placement coursing from the left upper quadrant into the left paracolic gutter. (b) Lateral X-ray demonstrating PD catheter placement coursing from the left upper quadrant into the left paracolic gutter. Photograph of PD catheter placement. Following catheter placement, manual PD using the Dialy-Nate® system was initiated with 16 mL volume dwells (∼20 mL/kg based on birth weight) of 45 min using 1.5% Dianeal® for 24 cycles per day. The dextrose concentration of the dialysate was later increased to 2.5% to optimize ultrafiltration. The infant developed hypotension, so dwell times were decreased to 30 min followed by 25-min drain periods, resulting in better blood pressures while maintaining adequate ultrafiltration and clearance. Within 72 h of initiating PD, the infant achieved a net ultrafiltration of 4–6 mL/pass. With successful dialysis, uremia improved; BUN decreased from 87 to 45 mg/dL. After 3 days, total fluid intake was liberalized to 120 mL/kg/day for venous nutrition. Dietary protein was liberalized to 3.5 g per kilogram per day. Enteral feedings were started on DOL 8. Weight decreased to 1024 g from a peak of 1092 g over the course of 10 days. The infant was transitioned from high frequency to conventional ventilation on DOL 13. By DOL 15, urine output had improved to 3–4 mL/kg/hr, and the infant maintained appropriate fluid balance during trials off PD. PD was discontinued on DOL 21. The catheter was removed on DOL 28. Adequate urine output was maintained during the remainder of his hospital admission without diuretic. Renal labs improved prior to discharge to a BUN of 28 mg/dL and a serum creatinine of 0.9 mg/dL. No acute infectious complications were experienced during the use of PD. There were no episodes of PD catheter leak. The infant was discharged to home on DOL 105. He is being followed by pediatric nephrology for the monitoring of his chronic kidney disease. He does not require RST or diuretic therapy. Current laboratory studies demonstrate a serum creatinine of 0.5 mg/dL at age 16 months, suggesting an estimated glomerular filtration rate of 61 mL/min/1.73 m2 by the modified Schwartz equation. He no longer requires supplemental oxygen or other respiratory support.

Discussion

To our knowledge, our case report includes the smallest infant with whom PD was successful using a standard, commercially available PD catheter. RST was indicated given severe oligo-anuric AKI resistant to medical management, fluid overload resulting in respiratory and hemodynamic compromise and inability to provide adequate nutrition secondary to fluid restriction. Use of RST in neonates is often limited by lack of size-specific equipment. PD was chosen for RST due to the lack of appropriately sized vascular catheters for a patient of such small size. Due to our infant's extremely small size and severe fluid overload, novel PD catheter placement was required. His initial dialysis prescription also required adaptation with increased initial fill volumes of 20 mL/kg to overcome the comparatively large amount of dead space within dialysis tubing (a catheter volume of 0.4 mL and a connecting tube volume of 3.0 mL). Limited existing literature documents high mortality rates in ELBW infants treated with PD, though the high rates in some reports are likely related to the underlying medical conditions. For example, two case series using PD in neonates each reported >50% mortality but included many infants with inborn errors of metabolism [6, 7]. Continuous veno-venous hemodialysis is the preferred therapy for patients with this condition because of its superior efficacy [10]; inefficient clearance with PD is not adequate for rapid removal of ammonia in these cases. Similarly, one group reported a 75% mortality rate in the use of PD for infants with necrotizing enterocolitis [11], a condition associated with high mortality rates as well as one which would be considered a relative contraindication to the use of PD [12] because of the inflammation in the abdomen and high risk for infection. In addition to high mortality rates, the available literature describes high rates of complications such as catheter leak and peritonitis during the use of PD in small infants, likely related to the lack of size-specific equipment for these patients. Use of varying catheter types, both commercial and improvised, for PD in ELBW neonates has been described. Two larger case series of PD in neonates (BW range, 1–4 kg) report technical success using rigid catheters in the infra-umbilical position [6, 7]. The patients in these series were all larger than ours. Oyachi et al. describe successful use of a flexible 10-French Blake silicone drain in a 40-week term infant requiring PD [13]. Yu et al. detail novel use of a vascular catheter placed in the right upper quadrant; however, high rates of catheter leak, poor drainage, and kinking were noted [14]. Similarly, use of a large-bore intravenous cannula has been described; however, again this improvised device was prone to malfunction with leak and kinking [9]. To our knowledge, this case illustrates the first time a standard, temporary PD catheter has been placed successfully via a left upper abdominal quadrant insertion site facilitating uncoiled extension into the left paracolic gutter in an ELBW infant. Early initiation of RST for AKI and fluid overload in neonates may have important short- and long-term systemic effects [15]. Adult and experimental data suggest that mortality due to AKI may be related to multi-organ functional and transcriptional changes, especially in the lungs [16-19]. AKI is implicated in the development of bronchopulmonary dysplasia due to pulmonary edema from volume overload and the associated elevation in pro-inflammatory cytokines [20]. The adult critical care literature demonstrates that critically ill patients with AKI are more difficult to wean from mechanical ventilation [21]. These findings bolster the need for early fluid removal and treatment of AKI in neonates given the implications for sequelae of chronic lung disease in this vulnerable population. Herein, the neonatology team was able to transition our infant from high frequency to conventional ventilation within 1 week of dialysis initiation. Our infant's severely fluid restricted state prior to RST initiation eliminated any opportunity to provide nutrition without further fluid accumulation. Adequate nutritional support may be unintentionally deferred in critically ill patients due to fluid restriction, preparation for surgical/procedural interventions and/or concern for feeding intolerance [22]. The resultant malnutrition in the setting of AKI is associated with increased morbidity and mortality [23, 24]. Nutritional deficit, specifically protein debt, is linked to higher risk for infection, reduced wound healing and increased need for mechanical ventilation [25]. Provision of adequate nutritional support to critically ill patients is a major benefit of RST.

Conclusion

The smallest of patients are at high risk for morbidity and mortality secondary to AKI, yet there remains a dearth of clinical equipment and expertise available to provide adequate RST for these patients. Technical adaptation is required for placement of PD catheters in ELBW infants; however, renal recovery following severe oligo-anuric AKI is possible and can be followed by excellent growth, development and improvement in lung function. These infants should not be excluded from consideration for initiation of RST because of their small size.
  24 in total

1.  Can peritoneal dialysis be used in preterm infants with gastrointestinal perforation?

Authors:  Fuat Emre Canpolat; Murat Yurdakök; Sule Yiğit; Gulsevin Tekinalp
Journal:  Pediatr Int       Date:  2010-10       Impact factor: 1.524

2.  Acute kidney injury in congenital diaphragmatic hernia requiring extracorporeal life support: an insidious problem.

Authors:  Samir K Gadepalli; David T Selewski; Robert A Drongowski; George B Mychaliska
Journal:  J Pediatr Surg       Date:  2011-04       Impact factor: 2.545

Review 3.  Management and outcomes of very low birth weight.

Authors:  Eric C Eichenwald; Ann R Stark
Journal:  N Engl J Med       Date:  2008-04-17       Impact factor: 91.245

Review 4.  Metabolic and nutritional aspects of acute renal failure in critically ill patients requiring continuous renal replacement therapy.

Authors:  Jennifer A Wooley; Imad F Btaiche; Kelley L Good
Journal:  Nutr Clin Pract       Date:  2005-04       Impact factor: 3.080

5.  The implementation of neonatal peritoneal dialysis in a clinical setting.

Authors:  Sevim Unal; Leyla Bilgin; Mehmet Gunduz; Nermin Uncu; Mujdem Nur Azili; Tugrul Tiryaki
Journal:  J Matern Fetal Neonatal Med       Date:  2012-03-16

6.  Acute kidney injury in asphyxiated newborns treated with therapeutic hypothermia.

Authors:  David T Selewski; Brian K Jordan; David J Askenazi; Ronald E Dechert; Subrata Sarkar
Journal:  J Pediatr       Date:  2012-11-10       Impact factor: 4.406

7.  Acute kidney injury reduces survival in very low birth weight infants.

Authors:  Rajesh Koralkar; Namasivayam Ambalavanan; Emily B Levitan; Gerald McGwin; Stuart Goldstein; David Askenazi
Journal:  Pediatr Res       Date:  2011-04       Impact factor: 3.756

8.  Acute kidney injury is independently associated with mortality in very low birthweight infants: a matched case-control analysis.

Authors:  David J Askenazi; Russell Griffin; Gerald McGwin; Waldemar Carlo; Namasivayam Ambalavanan
Journal:  Pediatr Nephrol       Date:  2009-02-24       Impact factor: 3.714

9.  Temporary peritoneal dialysis in newborns and children: a single-center experience over five years.

Authors:  Gurkan Genc; Unal Bicakci; Mithat Gunaydin; Burak Tander; Canan Aygun; Ozan Ozkaya; Riza Rizalar; Ender Ariturk; Sukru Kucukoduk; Ferit Bernay
Journal:  Ren Fail       Date:  2012-08-20       Impact factor: 2.606

10.  Peritoneal Dialysis in Neonates with Extremely Low Body Weight at Birth: New Modality of Using IV Cannula for Peritoneal Access.

Authors:  Vesna Stojanovic; Svetlana Bukarica; Aleksandra Doronjski; Smiljana Marinkovic
Journal:  Iran J Pediatr       Date:  2013-12       Impact factor: 0.364

View more
  12 in total

1.  Fluid overload and outcomes in neonates receiving continuous renal replacement therapy.

Authors:  Sang Taek Lee; Heeyeon Cho
Journal:  Pediatr Nephrol       Date:  2016-03-14       Impact factor: 3.714

2.  End-stage kidney disease in infancy: an educational review.

Authors:  Keia R Sanderson; Bradley A Warady
Journal:  Pediatr Nephrol       Date:  2018-11-21       Impact factor: 3.714

3.  Improving acute peritoneal dialysis outcome with use of soft peritoneal dialysis catheter (Cook Mac-Loc Multipurpose Drainage catheter®) among infants < 1500 g in a low resource setting.

Authors:  Rajiv Sinha; Rana Saha; Deblina Dasgupta; Niladri Bose; Shamik Ghosh; Ashok Modi; Bikramjit Das; Mignon McCulloch; Yincent Tse
Journal:  Pediatr Nephrol       Date:  2022-08-04       Impact factor: 3.651

4.  Kidney support for babies: building a comprehensive and integrated neonatal kidney support therapy program.

Authors:  Tahagod H Mohamed; Jolyn Morgan; Theresa A Mottes; David Askenazi; Jennifer G Jetton; Shina Menon
Journal:  Pediatr Nephrol       Date:  2022-10-13       Impact factor: 3.651

Review 5.  Neonatal acute kidney injury: a case-based approach.

Authors:  Michelle C Starr; Shina Menon
Journal:  Pediatr Nephrol       Date:  2021-02-17       Impact factor: 3.714

6.  Acute peritoneal dialysis is an efficient and reliable alternative therapy in preterm neonates with acute kidney injury.

Authors:  Yan Xing; Kai Sheng; Hui Liu; Song Wu; Hongling Wei; Rui Li; Jing Wang; Zailing Li; Xiaomei Tong
Journal:  Transl Pediatr       Date:  2021-04

Review 7.  Renal consequences of preterm birth.

Authors:  Amelie Stritzke; Sumesh Thomas; Harish Amin; Christoph Fusch; Abhay Lodha
Journal:  Mol Cell Pediatr       Date:  2017-01-18

8.  Successful peritoneal dialysis using a percutaneous tube for peritoneal drainage in an extremely low birth weight infant: a case report.

Authors:  Satoshi Yokoyama; Takayuki Nukada; Yuka Ikeda; Shigeto Hara; Akira Yoshida
Journal:  Surg Case Rep       Date:  2017-11-09

Review 9.  Renal replacement therapies for infants and children in the ICU.

Authors:  Keia R Sanderson; Lyndsay A Harshman
Journal:  Curr Opin Pediatr       Date:  2020-06       Impact factor: 2.893

10.  Efficacy and safety of mycophenolate mofetil in patients with IgA nephropathy: an update meta-analysis.

Authors:  Bing Du; Ye Jia; Wenhua Zhou; Xu Min; Lining Miao; Wenpeng Cui
Journal:  BMC Nephrol       Date:  2017-07-19       Impact factor: 2.388

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.