Feasibility of ASL spinal bone marrow perfusion imaging with optimized inversion time.

Dong Xing1, Yunfei Zha1, Liyong Yan1, Kejun Wang1, Wei Gong1, Hui Lin2.   

Abstract

PURPOSE: To assess the correlation between flow-sensitive alternating inversion recovery (FAIR) and dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) in the measurement of spinal bone marrow (SBM) perfusion; in addition, to assess for an optimized inversion time (TI) as well as the reproducibility of SBM FAIR perfusion.
MATERIALS AND METHODS: The optimized TI of a FAIR SBM perfusion experiment was carried out on 14 volunteers; two adjacent vertebral bodies were selected from each volunteer to measure the change of signal intensity (ΔM) and the signal-to-noise ratio (SNR) of FAIR perfusion MRI with five different TIs. Then, reproducibility of FAIR data from 10 volunteers was assessed by the reposition SBM FAIR experiments. Finally, FAIR and DCE-MRI were performed on 27 subjects. The correlation between the blood flow on FAIR (BFASL ) and perfusion-related parameters on DCE-MRI was evaluated.
RESULTS: The maximum value of ΔM and SNR were 36.39 ± 12.53 and 2.38 ± 0.97, respectively; both were obtained when TI was near 1200 msec. There were no significant difference between the two successive measurements of SBM BFASL perfusion (P = 0.879), and the within-subject coefficients of variation (wCV) of the measurements was 3.28%. The BFASL showed a close correlation with K(trans) (P < 0.001) and Kep (P = 0.004), and no correlation with Ve (P = 0.082) was found.
CONCLUSION: 1200 msec was the optimal TI for the SBM ASL perfusion image, which led to the maximum ΔM and a good quality perfusion image. The SBM FAIR perfusion scan protocol has good reproducibility, and as blood flow measurement on FAIR is reliable and closely related with the parameters on DCE-MRI, FAIR is feasible for measuring SBM blood flow.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  DCE-MRI; arterial spin labeling; magnetic resonance imaging; perfusion; spinal bone marrow

Mesh:

Substances:

Year:  2015        PMID: 25854511     DOI: 10.1002/jmri.24891

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  7T bone perfusion imaging of the knee using arterial spin labeling MRI.

Authors:  Xiufeng Li; Casey P Johnson; Jutta Ellermann
Journal:  Magn Reson Med       Date:  2019-12-24       Impact factor: 4.668

2.  [Test-retest reliability of 3D spiral fast-spin-echo pseudo-continuous arterial spin labeling for cerebral perfusion imaging of subcortical gray matter in healthy adults].

Authors:  Meng-Qi Liu; Zhi-Ye Chen; Lin Ma
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2017-09-20

3.  Arterial spin labeling of nasopharyngeal carcinoma shows early therapy response.

Authors:  Jun Liu; Juan Zhu; Yaxian Wang; Fei Wang; Hualin Yang; Nan Wang; Qingyun Chu; Qing Yang
Journal:  Insights Imaging       Date:  2022-07-07
  3 in total