E Babikir1, Hussein A Hasan2, A Abdelrazig3, M A Alkhorayef4, E Manssor5, A Sulieman5.
Abstract
This study aimed to assess patient entrance surface air kerma (ESAK) during chest and abdominal X-ray procedures in screen film radiography (SFR) and computed radiography (CR) to establish dose reference levels. Patients' doses were measured in five hospitals for a total of 196 patients. ESAK was calculated from exposure parameters using DosCal software. The X-ray tube output (mGy mAs(-1)), accuracy of exposure factors, linearity and reproducibility were measured using an Unfors Xi dosimeter. The overall mean and range of ESAK during chest X-ray were 0.6 ± 0.3 (0.1-1.3) mGy, while for abdominal X-rays they were 4.0 ± 3.2 (1.3-9.2) mGy. Hospital with a CR system was found to use relatively higher doses. Dose values for abdominal X-ray procedures were comparable with previous studies. The dose for chest X-ray procedure was higher by a factor of 2-3 compared with the current international reference levels.Entities:
Mesh:
Year: 2015 PMID: 25852182 DOI: 10.1093/rpd/ncv108
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972