Factors influencing delivered mean airway pressure during nasal CPAP with the RAM cannula.

Jeffrey S Gerdes1,2,3, Emidio M Sivieri1,2, Soraya Abbasi1,2,3.   

Abstract

OBJECTIVE: To measure mean airway pressure (MAP) delivered through the RAM Cannula® when used with a ventilator in CPAP mode as a function of percent nares occlusion in a simulated nasal interface/test lung model and to compare the results to MAPs using a nasal continuous positive airway pressure (NCPAP) interface with nares fully occluded. STUDY
DESIGN: An artificial airway model was connected to a spontaneous breathing lung model in which MAP was measured at set NCPAP levels between 4 and 8 cmH2 O provided by a Dräger Evita XL® ventilator and delivered through three sizes of RAM cannulae. Measurements were performed with varying leakage at the nasal interface by decreasing occlusion from 100% to 29%, half-way prong insertion, and simulated mouth leakage. Comparison measurements were made using the Dräger BabyFlow® NCPAP interface with a full nasal seal.
RESULTS: With simulated mouth closed, the Dräger interface delivered MAPs within 0.5 cmH2 O of set CPAP levels. For the RAM cannula, with 60-80% nares occlusion, overall delivered MAPs were 60 ± 17% less than set CPAP levels (P < 0.001). Further, MAP decreased progressively with decreasing percent nares occlusion. The simulated open mouth condition resulted in significantly lower MAPs to <1.7 cmH2 O. The one-half prong insertion depth condition, with closed mouth, yielded MAPs approximately 35 ± 9% less than full insertion pressures (P < 0.001).
CONCLUSIONS: In our bench tests, the RAM interface connected to a ventilator in NCPAP mode failed to deliver set CPAP levels when applied using the manufacturer recommended 60-80% nares occlusion, even with closed mouth and full nasal prong insertion conditions.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  NCPAP; mean airway pressure; nasal cannula; nasal prongs; premature infant; upper airway model

Mesh:

Year:  2015        PMID: 25851534     DOI: 10.1002/ppul.23197

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  8 in total

1.  Continuous positive airway pressure (CPAP) vs noninvasive positive pressure ventilation (NIPPV) vs noninvasive high frequency oscillation ventilation (NHFOV) as post-extubation support in preterm neonates: protocol for an assessor-blinded, multicenter, randomized controlled trial.

Authors:  Yuan Shi; Daniele De Luca
Journal:  BMC Pediatr       Date:  2019-07-26       Impact factor: 2.125

2.  RAM cannula versus short binasal prongs for nasal continuous positive airway pressure delivery in preterm infants: a randomized, noninferiority trial from low-middle-income country.

Authors:  S K Samim; Pradeep Kumar Debata; Anita Yadav; Jogender Kumar; Pratima Anand; Mehak Garg
Journal:  Eur J Pediatr       Date:  2022-09-17       Impact factor: 3.860

3.  Nonclinical Bench Performance Testing of a Very Low-Cost Nonelectric Bubble Continuous Positive Airway Pressure (bCPAP) and Blenders Device Designed for Newborn Respiratory Support.

Authors:  Patricia S Coffey; Alec Wollen
Journal:  Med Devices (Auckl)       Date:  2022-06-27

Review 4.  Preventing Continuous Positive Airway Pressure Failure: Evidence-Based and Physiologically Sound Practices from Delivery Room to the Neonatal Intensive Care Unit.

Authors:  Clyde J Wright; Laurie G Sherlock; Rakesh Sahni; Richard A Polin
Journal:  Clin Perinatol       Date:  2018-02-28       Impact factor: 3.430

Review 5.  Non-invasive Respiratory Support of the Premature Neonate: From Physics to Bench to Practice.

Authors:  Ibrahim Sammour; Sreenivas Karnati
Journal:  Front Pediatr       Date:  2020-05-08       Impact factor: 3.418

Review 6.  Pediatric pulmonology year in review 2016: Part 2.

Authors:  Richard Auten; Clement Ren; Ozge Yilmaz; Terry L Noah
Journal:  Pediatr Pulmonol       Date:  2017-04-25

Review 7.  Aerosol drug delivery to spontaneously-breathing preterm neonates: lessons learned.

Authors:  Federico Bianco; Fabrizio Salomone; Ilaria Milesi; Xabier Murgia; Sauro Bonelli; Elena Pasini; Raffaele Dellacà; Maria Luisa Ventura; Jane Pillow
Journal:  Respir Res       Date:  2021-02-26

8.  RAM cannula with Cannulaide versus Hudson prongs for delivery of nasal continuous positive airway pressure in preterm infants: an RCT.

Authors:  Shravani Maram; Srinivas Murki; Sidharth Nayyar; Sandeep Kadam; Tejo Pratap Oleti; Rajendra Prasad Anne; Saikiran Deshobhotla; Deepak Sharma; Subhash Arun; Praveen Rao Vadije
Journal:  Sci Rep       Date:  2021-12-07       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.