| Literature DB >> 25838938 |
Mehravar Rafati1, Masoud Arabfard2, Mehrdad Rafati Rahimzadeh3, Hasan Voshtani4, Hassan Moladoust4.
Abstract
BACKGROUND: Ultrasonic evaluation of intima-media thickness (IMT) is an early marker of assessing the development of atherosclerosis and determining cardiovascular risk. To attain the best possible diagnosis, it is essential that medical images be clear, sharp and without noise and artifacts.Entities:
Keywords: Brachial Artery; Carotid Intima-Media Thickness; Noise; Ultrasonography
Year: 2015 PMID: 25838938 PMCID: PMC4376986 DOI: 10.5812/ircmj.25013
Source DB: PubMed Journal: Iran Red Crescent Med J ISSN: 2074-1804 Impact factor: 0.611
Performance Metrics and Their Mathematical Definitions
| Performance | Mathematical Expression | Purpose |
|---|---|---|
|
| The MSE is the average of the square of the difference between original and de-noised image intensity values divided by the size of the image ( | |
|
| It is a ratio between the maximum possible power of the signal and the noise content. Higher PSNR value shows better image quality ( | |
|
| It measures the error perdition accuracy of the image. Large value of NAE indicates poor quality ( | |
|
| X | The closeness between two digital images can also be quantifies for correlation function. Normalized Cross Correlation measures the similarity between two images. Its value would approach 1 as difference between the two images approaches 0 ( |
|
| It is measure of average contrast of an image. A low SI value for effective de-speckling. Usually the speckle index changes from 1 to a lesser value ( | |
|
| It measures similarity between the original and de-noised images. Reconstructed image is of good quality if SC value lies near 1, greater values indicate poor quality image ( | |
|
| ENL = (NMV)2/(NSD)2, | To estimate the speckle noise level another assessment parameter known as ENL over a uniform region is used. A larger value of ENL shows a better quantitative performance ( |
|
| It is a combination of three factors such as loss of correlation, mean distortion and variance distortion. The range of image quality index is [-1 1]. For identical images, the value of images, the value of image quality index is unity ( | |
|
| MD = max|x(i,j) – y(i,j), Where x (i, j) represents the original image and y (i, j) represents the de-noised image and i and j are the pixel position of the M × N image. | MD is the maximum of the error signal (difference between the reference signal and test image). Large value of MD means that the image is poor quality ( |
|
| AD is simply the average of difference between the reference signal and test image. A lower value of MD means that image is poor quality ( | |
|
| The SSIM is the best method to evaluate image quality. Two images would have good similarity if SSIM value is 1 and poor if it is lesser than 0.9 ( |
Figure 1.Representation of Topological Derivative Diagram
Figure 2.Left Common Carotid and Brachial Arteries Ultrasound Image for One Sample
Mean and Standard Deviation of Image Quality Metrics for Three Filters With the Canny Edge Detection in Left Common Carotid and Brachial Arteries of 42 Study Subjects. [a,b,c]
| Filters | SRAD Carotid | DTD Carotid | CTD Carotid | SRAD Brachial | DTD Brachial | CTD Brachial |
|---|---|---|---|---|---|---|
|
| 0.00324108 ± 0.00015115 | 0.00008730 ± 0.00001541 | 0.00000473 ± 0.00000068 | 0.00383169 ± 0.00059373 | 0.00003584 ± 0.00000089 | 0.00000039 ± 0.00000011 |
|
| 73.01653278 ± 0.20225305 | 88.86324610 ± 0.81565367 | 95.42742459 ± 0.63852141 | 88.27062507 ± 0.62662068 | 92.57914974 ± 0.11065172 | 96.38840262 ± 1.24717083 |
|
| 0.02367264 ± 0.00065456 | 0.02758987 ± 0.00103658 | 0.00314536 ± 0.00044993 | 0.08993134 ± 0.00047839 | 0.03650177 ± 0.00045128 | 0.00052831 ± 0.00009334 |
|
| 0.93597502 ± 0.00190416 | 0.95926880 ± 0.01742364 | 0.98488556 ± 0.00002543 | 0.94851670 ± 0.00016193 | 0.95996759 ± 0.00001070 | 0.986584858 ± 0.01323544 |
|
| 0.91999900 ± 0.00000013 | 0.94997444 ± 0.00000121 | 0.97438041 ± 0.00058733 | 0.92996422 ± 0.00000130 | 0.94999964 ± 0.00000012 | 0.97960682 ± 0.00060962 |
|
| 0.00000497 ± 0.00000089 | 0.00000556 ± 0.00000096 | 0.00000544 ± 0.00000098 | 0.00000519 ± 0.00000091 | 0.00000700 ± 0.00000122 | 0.00000703 ± 0.00000122 |
|
| 1.07920751 ± 0.00211329 | 1.00013505 ± 0.00003092 | 0.99934641 ± 0.00018419 | 0.67309709 ± 0.00229816 | 1.00088738 ± 0.00030854 | 1.00004156 ± 0.00001575 |
|
| 0.46115492 ± 0.00881941 | 0.92080477 ± 0.00087706 | 0.95046500 ± 0.00105049 | 0.21272323 ± 0.00321600 | 0.92794113 ± 0.00059459 | 0.96941824 ± 0.00020984 |
|
| 0.62094131 ± 0.02801523 | 0.53425688 ± 0.04575904 | 0.16480626 ± 0.06260978 | 0.48859484 ± 0.05026766 | 0.39531867 ± 0.03183652 | 0.09944374 ± 0.03892293 |
|
| 0.03302151 ± 0.00112835 | 0.00371374 ± 0.00016375 | 0.00043881 ± 0.00006806 | 0.05708596 ± 0.00015745 | 0.00231214 ± 0.00002969 | 0.00003372 ± 0.00000587 |
|
| 0.48527880 ± 0.00539073 | 0.54529820 ± 0.00607544 | 0.59050902 ± 0.00813031 | 0.23697928 ± 0.00178620 | 0.30018565 ± 0.00046646 | 0.55891358 ± 0.00293672 |
aAbbreviations: MSE, mean squared error; PSNR, peak signal to noise ratio; NAE, normalized absolute error; NCC, normalized cross-correlation; SSIM, structure similarity image measures; SI, speckle index; SC, structural content; IQI, image quality index; MD, maximum difference; AD, average difference; ENL, equivalent number of looks.
bData are presented as Mean ± SD.
cAll P Values for ANOVA and Post Hoc Multiple Comparisons LSD Test Were < 0.001.