| Literature DB >> 25836472 |
Wei-Liang Hsu, Jeffrey Davis, Kaushik Balakrishnan, Mohammed Ibn-Elhaj, Shona Kroto, Neal Brock, Stanley Pau.
Abstract
This paper presents a polarization microscope using an infrared (IR) full-Stokes imaging polarimeter. The IR polarimeter utilizes an optimized interference-based micropolarizer design, and provides full-Stokes images with resolution of 1608 × 1208 at 35 frames/second. The device fabrication, instrument calibration, performance evaluation, and measurement results are presented. The measurement error of the imaging polarimeter is less than 3.5%, and the standard deviations are less than 2%.Year: 2015 PMID: 25836472 DOI: 10.1364/OE.23.004357
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894