A comparison of sequential and spiral scanning techniques in brain CT.

Ivana Pace, Francis Zarb.   

Abstract

PURPOSE: To evaluate and compare image quality and radiation dose of sequential computed tomography (CT) examinations of the brain and spiral CT examinations of the brain imaged on a GE HiSpeed NX/I Dual Slice 2CT scanner.
METHODS: A random sample of 40 patients referred for CT examination of the brain was selected and divided into 2 groups. Half of the patients were scanned using the sequential technique; the other half were scanned using the spiral technique. Radiation dose data—both the computed tomography dose index (CTDI) and the dose length product (DLP)—were recorded on a checklist at the end of each examination. Using the European Guidelines on Quality Criteria for Computed Tomography, 4 radiologists conducted a visual grading analysis and rated the level of visibility of 6 anatomical structures considered necessary to produce images of high quality.
RESULTS: The mean CTDI(vol) and DLP values were statistically significantly higher (P <.05) with the sequential scans (CTDI(vol): 22.06 mGy; DLP: 304.60 mGy • cm) than with the spiral scans (CTDI(vol): 14.94 mGy; DLP: 229.10 mGy • cm). The mean image quality rating scores for all criteria of the sequential scanning technique were statistically significantly higher (P <.05) in the visual grading analysis than those of the spiral scanning technique. DISCUSSION: In this local study, the sequential technique was preferred over the spiral technique for both overall image quality and differentiation between gray and white matter in brain CT scans. Other similar studies counter this finding.
CONCLUSION: The radiation dose seen with the sequential CT scanning technique was significantly higher than that seen with the spiral CT scanning technique. However, image quality with the sequential technique was statistically significantly superior (P <.05).

Entities:  

Mesh:

Year:  2015        PMID: 25835403

Source DB:  PubMed          Journal:  Radiol Technol        ISSN: 0033-8397


  4 in total

1.  Efficacy of breast shielding during head computed tomography examination.

Authors:  Nika Zalokar; Nejc Mekis
Journal:  Radiol Oncol       Date:  2020-08-01       Impact factor: 2.991

2.  Evaluation of Lower-Dose Spiral Head CT for Detection of Intracranial Findings Causing Neurologic Deficits.

Authors:  J G Fletcher; D R DeLone; A L Kotsenas; N G Campeau; V T Lehman; L Yu; S Leng; D R Holmes; P K Edwards; M P Johnson; G J Michalak; R E Carter; C H McCollough
Journal:  AJNR Am J Neuroradiol       Date:  2019-10-24       Impact factor: 3.825

3.  Ultra-low-dose sequential computed tomography for quantitative lung aeration assessment-a translational study.

Authors:  Lorenzo Ball; Anja Braune; Francesco Corradi; Claudia Brusasco; Alessandro Garlaschi; Thomas Kiss; Thomas Bluth; Francesca Simonassi; Alice Bergamaschi; Jörg Kotzerke; Marcus J Schultz; Marcelo Gama de Abreu; Paolo Pelosi
Journal:  Intensive Care Med Exp       Date:  2017-04-04

4.  Efficacy of breast shielding during head computed tomography examination.

Authors:  Nika Zalokar; Nejc Mekis
Journal:  Radiol Oncol       Date:  2020-08-01       Impact factor: 2.991

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.