Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques.

V Hernandez1, R Abella1, J F Calvo2, D Jurado-Bruggemann3, I Sancho4, P Carrasco5.   

Abstract

PURPOSE: Several authors have recommended a 2 mm tolerance for multileaf collimator (MLC) positioning in sliding window treatments. In volumetric modulated arc therapy (VMAT) treatments, however, the optimal tolerance for MLC positioning remains unknown. In this paper, the authors present the results of a multicenter study to determine the optimal tolerance for both techniques.
METHODS: The procedure used is based on dynalog file analysis. The study was carried out using seven Varian linear accelerators from five different centers. Dynalogs were collected from over 100,000 clinical treatments and in-house software was used to compute the number of tolerance faults as a function of the user-defined tolerance. Thus, the optimal value for this tolerance, defined as the lowest achievable value, was investigated.
RESULTS: Dynalog files accurately predict the number of tolerance faults as a function of the tolerance value, especially for low fault incidences. All MLCs behaved similarly and the Millennium120 and the HD120 models yielded comparable results. In sliding window techniques, the number of beams with an incidence of hold-offs >1% rapidly decreases for a tolerance of 1.5 mm. In VMAT techniques, the number of tolerance faults sharply drops for tolerances around 2 mm. For a tolerance of 2.5 mm, less than 0.1% of the VMAT arcs presented tolerance faults.
CONCLUSIONS: Dynalog analysis provides a feasible method for investigating the optimal tolerance for MLC positioning in dynamic fields. In sliding window treatments, the tolerance of 2 mm was found to be adequate, although it can be reduced to 1.5 mm. In VMAT treatments, the typically used 5 mm tolerance is excessively high. Instead, a tolerance of 2.5 mm is recommended.

Mesh:

Year:  2015        PMID: 25832081     DOI: 10.1118/1.4915541

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  The use of log file analysis within VMAT audits.

Authors:  Conor K McGarry; Christina E Agnew; Mohammad Hussein; Yatman Tsang; Alan R Hounsell; Catharine H Clark
Journal:  Br J Radiol       Date:  2016-04-13       Impact factor: 3.039

2.  Comparison of MLC positioning deviations using log files and establishment of specific assessment parameters for different accelerators with IMRT and VMAT.

Authors:  Xiutong Lin; Tao Sun; Xiao Liu; Guifang Zhang; Yong Yin
Journal:  Radiat Oncol       Date:  2022-07-16       Impact factor: 4.309

Review 3.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

4.  Comparison of complexity metrics for multi-institutional evaluations of treatment plans in radiotherapy.

Authors:  Victor Hernandez; Jordi Saez; Marlies Pasler; Diego Jurado-Bruggeman; Nuria Jornet
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-22
  4 in total