Spectra of clinical CT scanners using a portable Compton spectrometer.

H A Duisterwinkel1, J K van Abbema1, M J van Goethem2, R Kawachimaru1, L Paganini1, E R van der Graaf1, S Brandenburg1.   

Abstract

PURPOSE: Spectral information of the output of x-ray tubes in (dual source) computer tomography (CT) scanners can be used to improve the conversion of CT numbers to proton stopping power and can be used to advantage in CT scanner quality assurance. The purpose of this study is to design, validate, and apply a compact portable Compton spectrometer that was constructed to accurately measure x-ray spectra of CT scanners.
METHODS: In the design of the Compton spectrometer, the shielding materials were carefully chosen and positioned to reduce background by x-ray fluorescence from the materials used. The spectrum of Compton scattered x-rays alters from the original source spectrum due to various physical processes. Reconstruction of the original x-ray spectrum from the Compton scattered spectrum is based on Monte Carlo simulations of the processes involved. This reconstruction is validated by comparing directly and indirectly measured spectra of a mobile x-ray tube. The Compton spectrometer is assessed in a clinical setting by measuring x-ray spectra at various tube voltages of three different medical CT scanner x-ray tubes.
RESULTS: The directly and indirectly measured spectra are in good agreement (their ratio being 0.99) thereby validating the reconstruction method. The measured spectra of the medical CT scanners are consistent with theoretical spectra and spectra obtained from the x-ray tube manufacturer.
CONCLUSIONS: A Compton spectrometer has been successfully designed, constructed, validated, and applied in the measurement of x-ray spectra of CT scanners. These measurements show that our compact Compton spectrometer can be rapidly set-up using the alignment lasers of the CT scanner, thereby enabling its use in commissioning, troubleshooting, and, e.g., annual performance check-ups of CT scanners.

Entities:  

Mesh:

Year:  2015        PMID: 25832078     DOI: 10.1118/1.4915497

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Segmentation-free x-ray energy spectrum estimation for computed tomography using dual-energy material decomposition.

Authors:  Wei Zhao; Lei Xing; Qiude Zhang; Qingguo Xie; Tianye Niu
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-30

2.  A model-based scatter artifacts correction for cone beam CT.

Authors:  Wei Zhao; Don Vernekohl; Jun Zhu; Luyao Wang; Lei Xing
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.