| Literature DB >> 2583020 |
U Krause1, N Doetsch, P Meusers, H J Kock.
Abstract
A central venous catheter with a subcutaneous infusion port was implanted for chemotherapy into a 64-year-old woman with gastric lymphoma. Six weeks later, two weeks after a trivial trauma with hyperextension of the shoulder joint, it was found that the catheter had broken and its tip portion had embolized into the pulmonary artery: it was retrieved without difficulty via the femoral vein. This event suggests that pressure between clavicle and first rib can cause material fatigue in a silicon catheter. Such catheters should therefore not be implanted via the subclavian vein.Entities:
Mesh:
Substances:
Year: 1989 PMID: 2583020 DOI: 10.1055/s-2008-1066843
Source DB: PubMed Journal: Dtsch Med Wochenschr ISSN: 0012-0472 Impact factor: 0.628