BACKGROUND: A novel MRI-conditional robot that navigates focused ultrasound (FUS) for the treatment of gynaecological tumors endovaginally was developed.
METHODS: The robotic system has two PC-controlled axes (linear and angular). The robotic system was manufactured using a digital manufacturing 3D printer using acrylonitrile butadiene styrene (ABS) plastic. Evaluation of the device was performed in a 1.5T MRI using excised porcine tissue.
RESULTS: The robotic system was successfully tested for MRI safety and compatibility. The robotic system has been tested for its functionality for creating multiple (overlapping) lesions in an in vitro model.
CONCLUSIONS: An MRI-conditional FUS robotic system was developed that has the potential to create thermal lesions with the intention of treating gynaecological tumors. In the future a third axis will be needed that lifts the robot up or down in order to access vaginas which are at a variable height from the MRI table.
Copyright © 2015 John Wiley & Sons, Ltd.

Entities:  

Keywords:  MRI; gynaecological; robot

Mesh:

Year:  2015        PMID: 25808561     DOI: 10.1002/rcs.1653

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  2 in total

1.  Full coverage path planning algorithm for MRgFUS therapy.

Authors:  Anastasia Antoniou; Andreas Georgiou; Nikolas Evripidou; Christakis Damianou
Journal:  Int J Med Robot       Date:  2022-03-13       Impact factor: 2.483

2.  MRI-guided coupling for a focused ultrasound system using a top-to-bottom propagation.

Authors:  Marinos Yiannakou; George Menikou; Christos Yiallouras; Christakis Damianou
Journal:  J Ther Ultrasound       Date:  2017-01-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.