A comparison study between two analyzers for determining serum ferritin.

Xiaolian Zhang, Yu Lu, Liping Ma, Qiliu Peng, Xue Qin, Shan Li.   

Abstract

BACKGROUND: The measurement of serum ferritin (SF) is generally considered to be a reliable indicator of iron status in vivo. The accurate assessment of SF concentration promoted us to evaluate the analytical performance of two automated immunoassays.
METHODS: A total of 125 different concentrations of clinical samples were routinely collected and determined by two analytical systems: the Architect i2000 system from Abbott Laboratories and the Cobas E601 from Roche Diagnostics. We evaluated the analytical performance of the two systems and compared the measurements of SF results, which covered the whole analytical range.
RESULTS: The analytical performance and accuracy of the two systems were fairly good, and a satisfactory correlation was achieved between the two methods (r = 0.99), but Bland-Altman analysis suggested that the Cobas E601 was, on average, 60.6 ng/mL higher than the i2000, with wide limits of agreement (95% limits of agreement: -78.3 to 199.5 ng/mL).
CONCLUSIONS: These findings suggested that the two methods had good correlation but were not interchangeable. Patients should always be monitored for SF with the same method and it should be indicated in the report.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25807651     DOI: 10.7754/clin.lab.2014.140807

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  3 in total

1.  Measuring plasma ferritin levels with two different methods: A comparison of Roche Cobas e601 versus Roche Cobas c501 (integrated modular system Roche Cobas 6000).

Authors:  Durmuş Ayan; Sibel Soylemez
Journal:  J Med Biochem       Date:  2020-01-10       Impact factor: 3.402

2.  A Comparative Study for Measuring Serum Ferritin Levels with Three Different Laboratory Methods: Enzyme-Linked Immunosorbent Assay versus Cobas e411 and Cobas Integra 400 Methods.

Authors:  Lotfi S Bin Dahman; Khalid M Sumaily; Essa M Sabi; Mohammed A Hassan; Abeer M Bin Thalab; Asrar S Sayad; Saleh M Bin Kolaib; Fatima M Alhadhrmi
Journal:  Diagnostics (Basel)       Date:  2022-01-27

Review 3.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.