| Literature DB >> 25796040 |
Diming Zhang1, Jing Jiang2, Junye Chen3, Qian Zhang3, Yanli Lu3, Yao Yao3, Shuang Li3, Gang Logan Liu2, Qingjun Liu4.
Abstract
Rapid, sensitive, selective and portable detection of 2,4,6-trinitrotoluene (TNT) is in high demand for public safety and environmental monitoring. In this study, we reported a smartphone-based system using impedance monitoring for TNT detection. The screen-printed electrodes modified with TNT-specific peptides were used as disposable a biosensor to produce impedance responses to TNT. The responses could be monitored by a hand-held device and send out to smartphone through Bluetooth. Then, the smartphone was used to display TNT responses in real time and report concentration finally. In the measurement, the system was demonstrated to detect TNT at concentration as low as 10(-6) M and distinguish TNT versus different chemicals in high specificity. Thus, the smartphone-based biosensing platform provided a convenient and efficient approach to design portable instruments for chemical detections such as TNT recognition.Entities:
Keywords: 2 4 6-trinitrotoluene (TNT); Impedance; Peptide; Portable instrument; Screen-printed electrode; Smartphone
Mesh:
Substances:
Year: 2015 PMID: 25796040 DOI: 10.1016/j.bios.2015.03.004
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618