Breast dose reduction options during thoracic CT: influence of breast thickness.

Marie-Pierre Revel1, Isabelle Fitton, Etienne Audureau, Joseph Benzakoun, Mathieu Lederlin, Marie-Laure Chabi, Pascal Rousset.   

Abstract

OBJECTIVE: Little is known about the effectiveness of dose reduction options according to breast thickness. The purpose of this phantom study was to compare the effects on dose and noise of bismuth shielding versus a low kilovoltage for different breast thicknesses.
MATERIALS AND METHODS: CT acquisitions were performed first at 120 kVp (reference acquisition), then at 120 kVp with shielding and at 100 kVp without shielding on a phantom with three different prosthetic breast thicknesses, corresponding to the minimum, median, and maximum values first measured in a sample of 30 female thoracic CT examinations, which were randomly selected. Breast doses were measured with optically stimulated luminescence dosimeters placed on and beneath the prosthetic breast. For noise evaluation, the CT number SDs were measured within six ROIs at increasing depths.
RESULTS: Taking into account all breast thicknesses, the average breast dose was reduced by 42.1% with shielding and by 33.0% at 100 kVp (p=0.009). In-depth noise increased less with shielding (19.0% vs 32.1%, p<0.0001). For 1-cm breast thickness, the breast dose fell by 46.5% and 29.7% with shielding and 100 kVp, respectively (p=0.01), and in-depth noise increased by 19.5% and 33.9% (p=0.01). The corresponding values for 2-cm breast thickness were -38.5% and -30.1%, (p=0.02) and 16.5% and 33.5% (p=0.001), whereas those for 4-cm thickness were -40.6% and -40.5% (p=0.95) and 20.7 and 29.2% (p=0.02).
CONCLUSION: Greater breast dose reduction is achieved by shielding for breast thicknesses less than 4 cm. Regardless of breast thickness, shielding leads to a smaller increase in in-depth noise.

Entities:  

Keywords:  CT; breast dose; radiation dose; radiation savings

Mesh:

Substances:

Year:  2015        PMID: 25794091     DOI: 10.2214/AJR.14.13255

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  EANM guideline for ventilation/perfusion single-photon emission computed tomography (SPECT) for diagnosis of pulmonary embolism and beyond.

Authors:  Marika Bajc; Carl Schümichen; Thomas Grüning; Ari Lindqvist; Pierre-Yves Le Roux; Adriano Alatri; Ralf W Bauer; Mirza Dilic; Brian Neilly; Hein J Verberne; Roberto C Delgado Bolton; Bjorn Jonson
Journal:  Eur J Nucl Med Mol Imaging       Date:  2019-08-13       Impact factor: 9.236

2.  Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

Authors:  Ying-Lan Liao; Nan-Ku Lai; Yeu-Sheng Tyan; Hui-Yu Tsai
Journal:  Biomed J       Date:  2019-11-04       Impact factor: 4.910

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.