Schiotz tonometry accurately measures intraocular pressure in Boston type 1 keratoprosthesis eyes.

Igor E Estrovich1, Chris Shen, Yvonne Chu, J Crawford Downs, Stuart Gardiner, Michael Straiko, Steven L Mansberger.   

Abstract

PURPOSE: To evaluate the location of the eye and tonometry device that provides the most accurate intraocular pressure (IOP) measurements in eyes with Boston type 1 keratoprosthesis.
METHODS: A single surgeon sutured a Boston keratoprosthesis into the central cornea in the usual manner in 5 eyes of deceased donors. Another investigator used a 27-gauge needle, digital manometry, and gravity infusion to set the IOP at levels of 10, 20, 30, and 40 mm Hg in random order. A second investigator masked to the level of IOP used a Schiotz tonometer (Sklar) with a 7.5-g plunger load and a Tono-pen XL tonometer (Medtronic) to determine IOP at the temporal corneoscleral limbus and temporal sclera (3 mm temporal to the corneoscleral limbus). We used generalized estimation equation models to determine an average absolute difference between the tonometer measurements compared with the "gold standard" digital manometric IOP.
RESULTS: The Schiotz tonometer had a lower median absolute error compared with the Tono-pen at both temporal sclera (5.4 mm Hg vs. 39.0 mm Hg, P < 0.001) and corneoscleral limbus (4.9 vs. 15.0 mm Hg, P = 0.03). The median absolute error for Schiotz was similar at the temporal sclera and corneoscleral limbus (P = 0.44).
CONCLUSIONS: The Schiotz tonometer has the highest accuracy of IOP compared with the Tono-pen and gold standard digital manometry. Clinicians could consider Schiotz tonometry when measuring IOP in patients with keratoprosthesis.

Entities:  

Mesh:

Year:  2015        PMID: 25782403     DOI: 10.1097/ICO.0000000000000406

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  5 in total

1.  Effects of Medetomidine, Dexmedetomidine and their combination with Acepromazine on the intraocular pressure (IOP), tear secretion and pupil diameter in dogs.

Authors:  Ali Aghababaei; Ali Ronagh; Bahman Mosallanejad; Ali Baniadam
Journal:  Vet Med Sci       Date:  2021-03-10

2.  Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen.

Authors:  Di Ma; Chong-Bo Chen; Jiajian Liang; Zhihao Lu; Haoyu Chen; Mingzhi Zhang
Journal:  Sci Rep       Date:  2016-10-12       Impact factor: 4.379

Review 3.  Boston Type 1 Keratoprosthesis: Updated Perspectives.

Authors:  Manachai Nonpassopon; Muanploy Niparugs; Maria Soledad Cortina
Journal:  Clin Ophthalmol       Date:  2020-04-29

Review 4.  Challenges of Glaucoma Management in Patients with Type I Boston Keratoprosthesis.

Authors:  Sara M AlHilali; Samar A Al-Swailem
Journal:  Clin Ophthalmol       Date:  2022-02-11

5.  Comparison of Intraocular Pressure Measurement with Schiotz Tonometer and Tono-Pen Vet Tonometer in Healthy Dogs.

Authors:  Karolina Wrześniewska; Jacek Madany; Dagmara Winiarczyk
Journal:  J Vet Res       Date:  2018-10-24       Impact factor: 1.744

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.