Adequacy of powered vs manual bone marrow biopsy specimens: a retrospective review of sequential marrow aspirates and biopsies in 68 patients.

Douglas W Lynch1, Stephanie L Stauffer1, Nancy S Rosenthal2.   

Abstract

OBJECTIVES: To evaluate the quality and quantity of the bone marrow aspirates and biopsy specimens obtained with a powered system in comparison with the standard manual method.
METHODS: The Pathology Laboratory Information System was reviewed for patients who had previously undergone bone marrow biopsies performed by both the OnControl Bone Marrow System and the manual method. A total of 136 cases (68 patients) were reviewed for adequacy and compared using an unpaired t test.
RESULTS: The core biopsy specimens obtained by the OnControl system were significantly longer compared with those obtained by the manual system (16.9 vs 14.4 mm, P = .0036). However, the core biopsy specimens obtained by the manual method had on average more evaluable marrow elements (66% vs 40%, P < .0001), and the manual method was superior in 46 of the 68 cases when the length of evaluable marrow was calculated (9.7 vs 7 mm, P = .0049).
CONCLUSIONS: Our findings show that longer core biopsy specimens are obtained by the OnControl Bone Marrow system but that the manual method is still superior when the percentage and length of evaluable bone marrow are analyzed. Copyright© by the American Society for Clinical Pathology.

Entities:  

Keywords:  Bone marrow aspirate and biopsy; OnControl Bone Marrow System; Powered bone marrow system

Mesh:

Year:  2015        PMID: 25780005     DOI: 10.1309/AJCP67WITVPVVTNF

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  5 in total

1.  Quality and cost comparison of powered versus manual bone marrow biopsy devices in patients with myelofibrosis.

Authors:  Abhishek Maiti; Nicholas J Short; Srdan Verstovsek; Cynthia A Powers; Cheryl A Fullmer; Steven R Reyes; Carlos E Bueso-Ramos
Journal:  Leuk Lymphoma       Date:  2017-02-28

2.  CT-guided percutaneous biopsy of sclerotic bone lesions: diagnostic outcomes.

Authors:  I-Yuan Joseph Chang; Hakan Ilaslan; Murali Sundaram; Jean Schils; Naveen Subhas
Journal:  Skeletal Radiol       Date:  2017-12-07       Impact factor: 2.199

3.  Comparison of Bone Marrow Biopsy Specimens Obtained Using a Motorized Device and Manual Biopsy Systems.

Authors:  Catherine A Glennon; Janet M Woodroof; Suman Kambhampati; Alexis C Battershell; Serena R O'Connor; Kiley B Roberts
Journal:  Asia Pac J Oncol Nurs       Date:  2018 Oct-Dec

4.  Using Focused Laboratory Management and Quality Improvement Projects to Enhance Resident Training and Foster Scholarship.

Authors:  Matthew D Krasowski; Bradley A Ford; J Stacey Klutts; Chris S Jensen; Angela S Briggs; Robert A Robinson; Leslie A Bruch; Nitin J Karandikar
Journal:  Acad Pathol       Date:  2017-09-06

Review 5.  Practical Guide to Bone Marrow Sampling for Suspected Myelodysplastic Syndromes.

Authors:  Jean A Ridgeway; Sara Tinsley; Sandra E Kurtin
Journal:  J Adv Pract Oncol       Date:  2017-01-01
  5 in total