Simultaneous production of mixed electron--photon beam in a medical LINAC: A feasibility study.

Navid Khaledi1, Azim Arbabi2, Dariush Sardari3, Mohammad Mohammadi4, Ahmad Ameri2.   

Abstract

PURPOSE: The electron or photon beams might be used for treatment of tumors. Each beam has its own advantage and disadvantages. Combo beam can increase the advantages. No investigation has been performed for producing simultaneous mixed electron and photon beam. In current study a device has been added to the Medical Linac to produce a mixed photon-electron beam.
METHODS: Firstly a Varian 2300CD head was simulated by MCNP Monte Carlo Code. Two sets of perforated lead sheets with 1 and 2 mm thickness and 0.2, 0.3, and 0.5 cm punches then placed at the top of the applicator holder tray. This layer produces bremsstrahlung x-ray upon impinging fraction electrons on it. The remaining fraction of electrons passes through the holes. The simulation was performed for 10 × 10, 6 × 6, and 4 × 4 cm(2) field size.
RESULTS: For 10 × 10 cm(2) field size, among the punched targets, the largest penumbra difference between the depth of 1 and 7 cm was 72%. This difference for photon and electron beams were 31% and 325% respectively. A maximum of 39% photon percentage was produced by 2 mm target with 0.2 cm holes diameter layer. The minimum surface dose value was 4% lesser than pure electron beam. For small fields, unlike the pure electron beam, the PDD, penumbra, and flatness variations were negligible.
CONCLUSIONS: The advantages of mixing the electron and photon beam is reduction of pure electron's penumbra dependency with the depth, especially for small fields, also decreasing of dramatic changes of PDD curve with irradiation field size.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Medical LINAC; Mixed beam; Monte Carlo; Radiation therapy; Small fields

Mesh:

Year:  2015        PMID: 25773884     DOI: 10.1016/j.ejmp.2015.02.014

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Evaluation of Therapeutic Properties of a Low Energy Electron Beam Plus Spoiler for Local Treatment of Mycosis Fungoides: A Monte Carlo Study.

Authors:  Noshadi S; Atarod M; Amouheidari A; Felfeliyan F; Shokrani P
Journal:  J Biomed Phys Eng       Date:  2020-08-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.