Luis Condezo-Hoyos1, Elena Pérez-López1, Pilar Rupérez2.
Abstract
Optimization and validation of evaporative light scattering detector (ELSD), aided by response surface methodology (RSM), has been developed for the liquid chromatography analysis of a wide molecular weight (MW) range of carbohydrates, including polysaccharides and oligosaccharides. Optimal experimental parameters for the ELSD detection were: 88.8°C evaporator temperature, 77.9°C nebulizer temperature and 1.1 standard litres per minute nitrogen flow rate. Optimal ELSD detection, used together with high performance size exclusion chromatography (HPSEC) of carbohydrates, gave a linear range from 250 to 1000 mg L(-1) (R(2)>0.998), with limits of detection and quantitation of 4.83-11.67 and 16.11-38.91 mg L(-1), respectively. Relative standard deviation was lower than 1.8% for intra-day and inter-day repeatability for apple pectin, inulin, verbascose, stachyose and raffinose. Recovery ranged from 103.7% to 118.3% for fructo-oligosaccharides, α-galacto-oligosaccharides and disaccharides. Optimized and validated ELSD detection is proposed for the analysis of high- to low-MW carbohydrates with high sensitivity, precision and accuracy.Entities:
Keywords: Carbohydrate; ELSD; HPSEC; Oligosaccharide; Polysaccharide; RSM
Mesh:
Substances:
Year: 2015 PMID: 25766827 DOI: 10.1016/j.foodchem.2015.02.039
Source DB: PubMed Journal: Food Chem ISSN: 0308-8146 Impact factor: 7.514