Cardiac MRI in patients with cardiac pacemakers: practical methods for reducing susceptibility artifacts and optimizing image quality.

Touko Kaasalainen1, Sari Kivistö2, Miia Holmström2, Juha Peltonen3, Sami Pakarinen4, Helena Hänninen4, Outi Sipilä2.   

Abstract

BACKGROUND: Magnetic resonance imaging (MRI) of pacemaker patients has become available despite of previous contraindications. However, pacing systems containing ferromagnetic material may hamper the diagnostic quality of cardiac MR (CMR) images.
PURPOSE: To study methods for reducing susceptibility-based artifacts in CMR examinations of pacemaker patients.
MATERIAL AND METHODS: Altogether 16 patients were scanned with 1.5T MRI scanner using cine balanced steady-state free-precession (bSSFP) and spoiled gradient echo (SPGR) sequences. The use of frequency-scout was also evaluated. For myocardial late gadolinium-enhanced (LGE) imaging, SPGR or bSSFP readout inversion-recovery prepared gradient echo sequences were used with and without phase-sensitive inversion-recovery (PSIR). Two radiologists subjectively compared the image quality (IQ) and the ranges of susceptibility artifacts were evaluated objectively.
RESULTS: The IQ proved adequate for diagnosing each patient, although in a few patients with a left-side implanted generator, artifacts hampered IQ in the anterior and anteroseptal segments of the myocardium in bSSFP cine and LGE sequences. In bSSFP cine, the use of frequency-scout could often transfer the banding artifacts away from the left ventricular myocardium. In LGE imaging, the artifacts were more pronounced in IR-bSSFP and PSIR than in IR-SPGR sequences. The ranges of generator-based artifacts were greater in bSSFP (10-12 cm) than in SPGR (6 cm) sequences due to banding artifacts.
CONCLUSION: The artifacts caused by pacemakers typically did not compromise the diagnostic IQ. The use of frequency-scout prior to bSSFP cine or the use of SPGR-based sequences could also improve IQ. © The Foundation Acta Radiologica 2015.

Entities:  

Keywords:  CMR; frequency-scout; image quality; pacemaker; susceptibility artifact

Mesh:

Year:  2015        PMID: 25766728     DOI: 10.1177/0284185115574873

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  5 in total

Review 1.  Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5 T: influence of pacemaker related artifacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement, short tau inversion recovery and T1-weighted imaging.

Authors:  Oliver Klein-Wiele; Marietta Garmer; Martin Busch; Serban Mateiescu; Rhyan Urbien; Gianluca Barbone; Kaffer Kara; Michael Schulte-Hermes; Frauke Metz; Birgit Hailer; Dietrich Grönemeyer
Journal:  Int J Cardiovasc Imaging       Date:  2016-11-04       Impact factor: 2.357

Review 2.  Compact pediatric cardiac magnetic resonance imaging protocols.

Authors:  Evan J Zucker
Journal:  Pediatr Radiol       Date:  2022-07-12

3.  Reducing cardiac implantable electronic device-induced artefacts in cardiac magnetic resonance imaging.

Authors:  Aino-Maija Vuorinen; Lauri Lehmonen; Jarkko Karvonen; Miia Holmström; Sari Kivistö; Touko Kaasalainen
Journal:  Eur Radiol       Date:  2022-08-27       Impact factor: 7.034

4.  Cardiac Magnetic Resonance Imaging-Based Screening for Cardiac Sarcoidosis in Patients With Atrioventricular Block Requiring Temporary Pacing.

Authors:  Aino-Maija Vuorinen; Jukka Lehtonen; Sami Pakarinen; Miia Holmström; Sari Kivistö; Touko Kaasalainen
Journal:  J Am Heart Assoc       Date:  2022-06-06       Impact factor: 6.106

5.  Promoting the use of the PI-QUAL score for prostate MRI quality: results from the ESOR Nicholas Gourtsoyiannis teaching fellowship.

Authors:  Francesco Giganti; Alexander P Cole; Fiona M Fennessy; Timothy Clinton; Pedro Lopes Da Frota Moreira; Mariana Costa Bernardes; Carl-Fredrik Westin; Deepa Krishnaswamy; Andriy Fedorov; Daniel A Wollin; Bjoern Langbein; Nicola Frego; Muhieddine Labban; Joy S Badaoui; Steven L Chang; Logan G Briggs; Junichi Tokuda; Alessandro Ambrosi; Alex Kirkham; Mark Emberton; Veeru Kasivisvanathan; Caroline M Moore; Clare Allen; Clare M Tempany
Journal:  Eur Radiol       Date:  2022-06-30       Impact factor: 7.034

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.