Comparative dosimetry study of three UK centres implementing total skin electron treatment through external audit.

S Misson-Yates1, R Gonzalez, M McGovern, A Greener.   

Abstract

OBJECTIVE: This article describes the external audit measurements conducted in two UK centres implementing total skin electron beam therapy (TSEBT) and the results obtained.
METHODS: Measurements of output, energy, beam flatness and symmetry at a standard distance (95 or 100 cm SSD) were performed using a parallel plate chamber in solid water. Similarly, output and energy measurements were also performed at the treatment plane for single and dual fields. Clinical simulations were carried out using thermoluminescent dosemeters (TLDs) and Gafchromic® film (International Specialty Products, Wayne, NJ) on an anthropomorphic phantom.
RESULTS: Extended distance measurements confirmed that local values for the beam dosimetry at Centres A and B were within 2% for outputs and 1-mm agreement of the expected depth at which the dose is 50% of the maximum for the depth-dose curve in water (R50,D) value. Clinical simulation using TLDs) showed an agreement of -1.6% and -6.7% compared with the expected mean trunk dose for each centre, respectively, and a variation within 10% (±1 standard deviation) across the trunk. The film results confirmed that the delivery of the treatment technique at each audited centre complies with the European Organisation for Research and Treatment of Cancer recommendations.
CONCLUSION: This audit methodology has proven to be a successful way to confirm the agreement of dosimetric parameters for TSEBT treatments at both audited centres and could serve as the basis for an audit template to be used by other audit groups. ADVANCES IN KNOWLEDGE: TSEBT audits are not established in the UK owing to a limited number of centres carrying out the treatment technique. This article describes the audits performed at two UK centres prior to their clinical implementation.

Entities:  

Mesh:

Year:  2015        PMID: 25761213      PMCID: PMC4628479          DOI: 10.1259/bjr.20140723

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  6 in total

1.  The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration.

Authors:  D I Thwaites; A R DuSautoy; T Jordan; M R McEwen; A Nisbet; A E Nahum; W G Pitchford
Journal:  Phys Med Biol       Date:  2003-09-21       Impact factor: 3.609

2.  Analysis of regional radiotherapy dosimetry audit data and recommendations for future audits.

Authors:  A Palmer; B Mzenda; J Kearton; R Wills
Journal:  Br J Radiol       Date:  2010-12-15       Impact factor: 3.039

Review 3.  Clinical implementation of total skin electron beam (TSEB) therapy: a review of the relevant literature.

Authors:  S Diamantopoulos; K Platoni; M Dilvoi; I Nazos; K Geropantas; G Maravelis; M Tolia; I Beli; E Efstathopoulos; P Pantelakos; G Panayiotakis; V Kouloulias
Journal:  Phys Med       Date:  2010-10-14       Impact factor: 2.685

4.  A national dosimetry audit of intraoperative radiotherapy.

Authors:  D J Eaton; B Earner; P Faulkner; N Dancer
Journal:  Br J Radiol       Date:  2013-10-16       Impact factor: 3.039

Review 5.  Skin lymphoma.

Authors:  S L Morris
Journal:  Clin Oncol (R Coll Radiol)       Date:  2012-04-04       Impact factor: 4.126

6.  Total skin electron radiation in the management of mycosis fungoides: Consensus of the European Organization for Research and Treatment of Cancer (EORTC) Cutaneous Lymphoma Project Group.

Authors:  Glenn W Jones; Barry M Kacinski; Lynn D Wilson; Rein Willemze; Margaret Spittle; Gerda Hohenberg; Leonore Handl-Zeller; Franz Trautinger; Robert Knobler
Journal:  J Am Acad Dermatol       Date:  2002-09       Impact factor: 11.527

  6 in total
  2 in total

1.  Development of a novel program for conversion from tetrahedral-mesh-based phantoms to DICOM dataset for radiation treatment planning: TET2DICOM.

Authors:  Bo-Wi Cheon; Se Hyung Lee; Min Cheol Han; Chul Hee Min; Haegin Han; Chan Hyeong Kim; Jin Sung Kim
Journal:  J Appl Clin Med Phys       Date:  2021-10-11       Impact factor: 2.102

2.  Total skin electron beam therapy rationalization and utility of in vivo dosimetry in a high-volume centre.

Authors:  Sarah Misson-Yates; Marium Naeem; Isabel Palmer; Eleanor Holden; Owen Hedley; Mark McGovern; Stephen Morris; Anthony G Greener
Journal:  BJR Open       Date:  2019-07-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.