Literature DB >> 25760513

Modular hip implant fracture at the stem-sleeve interface.

Thomas Parisi, Brian Burroughs, Young-Min Kwon.   

Abstract

The use of modular implants in femoral stem design has grown increasingly popular over the last decade because of the theoretical advantage of more flexibility and optimization of femoral anteversion, limb length, and femoral component offset. With the benefit of increased surgical flexibility, however, modularity also carries the theoretical risks of fretting at the modular surfaces, sequelae of wear debris, and possible failure and fracture of the stem at the modular junction. Indeed, there have been an increasing number of reports of modular implants failing due to fracture at modular junctions. The S-ROM prosthesis (DePuy Orthopaedics, Inc, Warsaw, Indiana), however, has a stellar clinical record and has been used with good results in both primary and revision total hip arthroplasty. Only a single case of S-ROM failure at the stem-sleeve interface has been reported in the orthopedic literature. The aim of this case report was to present a succinct history of proximal modularity in total hip arthroplasty and to describe the only known case of this type of catastrophic failure in an S-ROM prosthesis with a metal-on-metal bearing. Despite a low level of serum metal ions on presentation, scanning electron microscopy showed findings consistent with corrosive processes and pseudotumor was seen at revision surgery. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25760513     DOI: 10.3928/01477447-20150305-91

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  10 in total

1.  Spontaneous Fractures of a Modern Modular Uncemented Femoral Stem.

Authors:  Chelsea N Koch; Laura Serrano Mateo; Stephen Kayiaros; Kevin A Cassidy; Seth A Jerabek; Alejandro Gonzalez Della Valle
Journal:  HSS J       Date:  2016-06-21

2.  Medium-term outcomes of the S-ROM modular femoral stem in revision hip replacement.

Authors:  Jesús Moreta; Iker Uriarte; Xabier Foruria; Ane Loroño; Urko Agirre; Iñaki Jáuregui; José Luis Martínez-de Los Mozos
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-04-25

3.  [Total hip arthroplasty with Wagner Cone stem and subtrochanteric shortening osteotomy in treatment of Crowe developmental dysplasia of hip in adults].

Authors:  Wei Lu; Min Zeng; Pengfei Lei; Jie Xie; Yihe Hu
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2019-08-15

4.  Modular prosthesis fracture in a patient with developmental dysplasia of the hip: a case report and literature review.

Authors:  Yuan-Pei Cheng; Xiao-Kang Cheng; Yong-Bo Li; Qian-Ru Zhang; Hao Feng; Yi-Han Zhong; Yan-Bo Zhang; Han Wu
Journal:  BMC Musculoskelet Disord       Date:  2021-05-14       Impact factor: 2.362

5.  Comparison of Clinical Efficacy Between Modular Cementless Stem Prostheses and Coated Cementless Long-Stem Prostheses on Bone Defect in Hip Revision Arthroplasty.

Authors:  Huibin Li; Fang Chen; Zhe Wang; Qian Chen
Journal:  Med Sci Monit       Date:  2016-02-29

6.  Developmental hip dysplasia treated by total hip arthroplasty using a cementless Wagner cone stem in young adult patients with a small physique.

Authors:  Ping Zhen; Jun Liu; Hao Lu; Hui Chen; Xusheng Li; Shenghu Zhou
Journal:  BMC Musculoskelet Disord       Date:  2017-05-15       Impact factor: 2.362

Review 7.  Modular Stems: Advantages and Current Role in Primary Total Hip Arthroplasty.

Authors:  Chan-Woo Park; Seung-Jae Lim; Youn-Soo Park
Journal:  Hip Pelvis       Date:  2018-09-04

8.  Fracture of an S-ROM stem at the sleeve-stem junction.

Authors:  Kilian Rueckl; Friedrich Boettner; Ulrich Bechler; Elexis C Baral; Timothy M Wright; Peter K Sculco
Journal:  Arthroplast Today       Date:  2018-05-10

9.  Fracture of a Sivash-Range of Motion Femoral Stem at the Diaphyseal Spline.

Authors:  Tyler J Humphrey; Daniel Marchwiany; Hany S Bedair; Christopher M Melnic
Journal:  Cureus       Date:  2021-12-24

10.  Femoral Stem Fracture in Hip Revision Arthroplasty: A Systematic Literature Review of the Real-World Evidence.

Authors:  Matthias Sukopp; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Z Orthop Unfall       Date:  2021-04-13       Impact factor: 0.923

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.