| Literature DB >> 25758294 |
Rui Zhang1, Yong Mei1, Yanru Liu1, Hao Dai1, Hongfang Xia1, Xin Zhang1, Yukang Wu1, Yingying Gu1, Xiaowu Peng2.
Abstract
The abundant production of methyl tert-butyl ether (MTBE) and its widespread use have led to an increase in the potential for human exposure. This work described a simple, fast, sensitive, reliable and low-cost method for the simultaneous measurement of MTBE and its metabolite, tert-butyl alcohol (TBA) in human serum by headspace solid-phase microextraction gas chromatography-mass spectrometry. Extraction conditions were optimized and 40 °C, 10 min, 250 rpm and 0.3 g NaCl for a 1 mL sample were the optimal conditions. This method showed good analytical performance in terms of sensitivity with limits of detection in serum (1 mL) of 0.03 µg/L for MTBE and 0.05 µg/L for TBA, accuracy (mean recovery values) from 75.8% to 85.8%, precision (relative standard deviations) <10% and sample stability (biodegradation) <10% after 28 days. A verification experiment proved the reproducibility and stability of this method as well. Finally the method was used to detect 212 specimens, and the internal dose levels for MTBE in human serum were presented in China.Entities:
Keywords: GC-MS; SPME; methyl tert-butyl ether; serum; tert-butyl alcohol
Mesh:
Substances:
Year: 2015 PMID: 25758294 DOI: 10.1002/bmc.3448
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902