Intraosseous infusion rates under high pressure: a cadaveric comparison of anatomic sites.

Jason Pasley1, Catriona H T Miller, Joseph J DuBose, Stacy A Shackelford, Raymond Fang, Kimberly Boswell, Chuck Halcome, Jonathan Casey, Michael Cotter, Michael Matsuura, Nathaniel Relph, Nicholas T Tarmey, Deborah M Stein.   

Abstract

BACKGROUND: When traditional vascular access methods fail, emergency access through the intraosseous (IO) route can be lifesaving. Fluids, medications, and blood components have all been delivered through these devices. We sought to compare the performance of IO devices placed in the sternum, humeral head, and proximal tibia using a fresh human cadaver model.
METHODS: Commercially available IO infusion devices were placed into fresh human cadavers: sternum (FAST-1), humeral head (EZ-IO), and proximal tibia (EZ-IO). Sequentially, the volume of 0.9% saline infused into each site under 300 mm Hg pressure over 5 minutes was measured. Rates of successful initial IO device placement and subjective observations related to the devices were also recorded.
RESULTS: For 16 cadavers over a 5-minute bolus infusion, the total volume of fluid infused at the three IO access sites was 469 (190) mL for the sternum, 286 (218) mL for the humerus, and 154 (94) mL for the tibia. Thus, the mean (SD) flow rate infused at each site was as follows: (1) sternum, 93.7 (37.9) mL/min; (2) humerus, 57.1 (43.5) mL/min; and (3) tibia, 30.7 (18.7) mL/min. The tibial site had the greatest number of insertion difficulties.
CONCLUSION: This is the first study comparing the rate of flow at the three most clinically used adult IO infusion sites in an adult human cadaver model. Our results showed that the sternal site for IO access provided the most consistent and highest flow rate compared with the humeral and tibial insertion sites. The average flow rate in the sternum was 1.6 times greater than in the humerus and 3.1 times greater than in the tibia.

Entities:  

Mesh:

Year:  2015        PMID: 25757113     DOI: 10.1097/TA.0000000000000516

Source DB:  PubMed          Journal:  J Trauma Acute Care Surg        ISSN: 2163-0755            Impact factor:   3.313


  12 in total

Review 1.  Intraosseous access in the resuscitation of trauma patients: a literature review.

Authors:  Joseph Antony Tyler; Zane Perkins; Henry Dudley De'Ath
Journal:  Eur J Trauma Emerg Surg       Date:  2020-02-20       Impact factor: 3.693

2.  Ventriculo-humeral shunt: a cadaveric feasibility study with application to treating hydrocephalus.

Authors:  Mohammad W Kassem; Basem Ishak; Karishma Mehta; Joe Iwanaga; Marios Loukas; R Shane Tubbs
Journal:  Childs Nerv Syst       Date:  2019-05-04       Impact factor: 1.475

Review 3.  Intramedullary placement of ventricular shunts: a review of using bone as a distal cerebrospinal absorption site in treating hydrocephalus.

Authors:  Mohammad W Kassem; Joshua Chern; Marios Loukas; R Shane Tubbs
Journal:  Childs Nerv Syst       Date:  2017-08-23       Impact factor: 1.475

4.  Bent Metal in a Bone: A Rare Complication of an Emergent Procedure or a Deficiency in Skill Set?

Authors:  Mridula Krishnan; Katherine Lester; Amber Johnson; Kaye Bardeloza; Peter Edemekong; Ilya Berim
Journal:  Case Rep Crit Care       Date:  2016-11-27

5.  Deformation of a humeral intraosseous catheter due to positioning for thoracostomy.

Authors:  Cliff Reid; Toby Fogg; Geoff Healy
Journal:  Clin Exp Emerg Med       Date:  2018-09-30

6.  Comparison of ROTEM parameters from venous and intraosseous blood.

Authors:  Marion Wiegele; Thomas Hamp; Johannes Gratz; Eleonore Pablik; Eva Schaden
Journal:  Sci Rep       Date:  2019-03-06       Impact factor: 4.379

7.  Intraosseous Catheter Flow Rates and Ease of Placement at Various Sites in Canine Cadavers.

Authors:  James Lange; Søren R Boysen; Adam Bentley; Aylin Atilla
Journal:  Front Vet Sci       Date:  2019-09-19

Review 8.  Sternal Intraosseous Devices: Review of the Literature.

Authors:  Jared A Laney; Jonathan Friedman; Andrew D Fisher
Journal:  West J Emerg Med       Date:  2021-03-24

9.  Comparison of the Fluid Resuscitation Rate with and without External Pressure Using Two Intraosseous Infusion Systems for Adult Emergencies, the CITRIN (Comparison of InTRaosseous infusion systems in emergency medicINe)-Study.

Authors:  Niels Hammer; Robert Möbius; André Gries; Björn Hossfeld; Ingo Bechmann; Michael Bernhard
Journal:  PLoS One       Date:  2015-12-02       Impact factor: 3.240

Review 10.  Use of intra-osseous access in adults: a systematic review.

Authors:  F Petitpas; J Guenezan; T Vendeuvre; M Scepi; D Oriot; O Mimoz
Journal:  Crit Care       Date:  2016-04-14       Impact factor: 9.097