Literature DB >> 25756936

Hemodynamic comparison between Trifecta and freestyle aortic valve during exercise in patients with small aortic root.

Giordano Tasca1, Paola Redaelli, Beatrice Riva, Caterina Chiara De Carlini, Elisabetta Lobiati, Amando Gamba.   

Abstract

BACKGROUND: Patients with a small aortic annulus, that is ≤ 23 mm, constitute a challenge for the surgeon, because they are at high risk of patient-prosthesis mismatch. Stentless valves provide better hemodynamic performance at rest and during exercise than stented valves, and are advocated in this group of patients. A new-generation stented valve, the Trifecta (St. Jude), has recently become available with improved hemodynamics. The aim of this study was to compare the hemodynamic performance of Freestyle (Medtronic) and Trifecta at rest and during exercise in patients with a small aortic annulus.
METHODS: From September 2012 to September 2014, 22 patients with a native aortic annulus ≤ 23 mm underwent ergometric stress testing one year after aortic valve replacement with either a Trifecta (12 patients) or a Freestyle (10 patients) bioprosthesis as part of a randomized study.
RESULTS: The mean gradient at rest was 6.0 ± 2.3 mmHg for Trifecta and 4.3 ± 3.5 for Freestyle (p = 0.213). The mean gradient at peak of exercise was 9.7 ± 3.4 mmHg for Trifecta and 7.4 ± 5 mmHg for Freestyle (p = 0.243). No significant differences were found between the two prostheses regarding other hemodynamic parameters: effective orifice area, velocity index, and performance indexes.
CONCLUSION: Both the stented Trifecta and stentless Freestyle prostheses provide excellent hemodynamic results during physical stress in patients with a small aortic annulus. Our study confirms that Trifecta implantation results in low gradients at rest and during exercise and that the performance of Trifecta is similar to that of a stentless valve.
© 2015 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2015        PMID: 25756936     DOI: 10.1111/jocs.12536

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  2 in total

1.  Implantation of Perceval in Trifecta ring: A new perspective.

Authors:  Tamer Owais; Mohammad El Garhy; Maria-Anna Secknus; Thomas Kuntze
Journal:  J Cardiol Cases       Date:  2019-08-22

2.  Early Hemodynamics after Aortic Valve Replacement.

Authors:  Serik Aitaliyev; Egle Rumbinaitė; Karolina Mėlinytė-Ankudavičė; Rokas Nekrošius; Vytenis Keturakis; Rimantas Benetis
Journal:  Medicina (Kaunas)       Date:  2020-12-07       Impact factor: 2.430

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.