MRI-based computed tomography metal artifact correction method for improving proton range calculation accuracy.

Peter C Park1, Eduard Schreibmann2, Justin Roper2, Eric Elder2, Ian Crocker2, Tim Fox3, X Ronald Zhu1, Lei Dong4, Anees Dhabaan5.   

Abstract

PURPOSE: Computed tomography (CT) artifacts can severely degrade dose calculation accuracy in proton therapy. Prompted by the recently increased popularity of magnetic resonance imaging (MRI) in the radiation therapy clinic, we developed an MRI-based CT artifact correction method for improving the accuracy of proton range calculations. METHODS AND MATERIALS: The proposed method replaces corrupted CT data by mapping CT Hounsfield units (HU number) from a nearby artifact-free slice, using a coregistered MRI. MRI and CT volumetric images were registered with use of 3-dimensional (3D) deformable image registration (DIR). The registration was fine-tuned on a slice-by-slice basis by using 2D DIR. Based on the intensity of paired MRI pixel values and HU from an artifact-free slice, we performed a comprehensive analysis to predict the correct HU for the corrupted region. For a proof-of-concept validation, metal artifacts were simulated on a reference data set. Proton range was calculated using reference, artifactual, and corrected images to quantify the reduction in proton range error. The correction method was applied to 4 unique clinical cases.
RESULTS: The correction method resulted in substantial artifact reduction, both quantitatively and qualitatively. On respective simulated brain and head and neck CT images, the mean error was reduced from 495 and 370 HU to 108 and 92 HU after correction. Correspondingly, the absolute mean proton range errors of 2.4 cm and 1.7 cm were reduced to less than 2 mm in both cases.
CONCLUSIONS: Our MRI-based CT artifact correction method can improve CT image quality and proton range calculation accuracy for patients with severe CT artifacts.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25752400     DOI: 10.1016/j.ijrobp.2014.12.027

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  1 in total

1.  Development of a hybrid magnetic resonance/computed tomography-compatible phantom for magnetic resonance guided radiotherapy.

Authors:  Min-Joo Kim; Seu-Ran Lee; Kyu-Ho Song; Hyeon-Man Baek; Bo-Young Choe; Tae Suk Suh
Journal:  J Radiat Res       Date:  2020-03-23       Impact factor: 2.724

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.