Literature DB >> 25749607

Compensation of matrix effects in a standard addition method for metformin in postmortem blood using liquid chromatography-electrospray-tandem mass spectrometry.

Merja Gergov1, Timo Nenonen1, Ilkka Ojanperä1, Raimo A Ketola2.   

Abstract

This work describes a procedure to evaluate matrix effects in a combined dilution and standard addition method (SAM) using liquid chromatography-electrospray-tandem mass spectrometry. The method was validated and applied to an analysis of metformin in postmortem blood samples. The analytical method included protein precipitation with methanol, followed by liquid chromatographic separation of metformin on Gemini NX-C18 reversed-phase column using a gradient consisting of methanol and ammonium acetate at pH 3.2. The mass spectrometric analysis was performed with a quadrupole-linear ion trap mass spectrometer equipped with a turbo ion spray interface in a positive ion mode using selected reaction monitoring. Quantitation was performed based on an SAM. Validation for metformin revealed a practical limit of quantification of 0.1 mg/L, a linear range from 0.1 to 3.0 mg/L, average precision 10%, accuracy (bias) 9% and reproducibility 10%. Combined matrix effects were evaluated by k-values (slopes) of calibration plots, postextraction addition approach and a comparison of within- and between-sample precision (relative standard deviation). It was demonstrated that the method contained matrix effects which were fully compensated for using dilution and the SAM.
© The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25749607     DOI: 10.1093/jat/bkv020

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  5 in total

1.  Post-mortem analysis of lactate concentration in diabetics and metformin poisonings.

Authors:  T Keltanen; T Nenonen; R A Ketola; I Ojanperä; A Sajantila; K Lindroos
Journal:  Int J Legal Med       Date:  2015-09-01       Impact factor: 2.686

2.  Determination of metformin bio-distribution by LC-MS/MS in mice treated with a clinically relevant paradigm.

Authors:  Kiran Chaudhari; Jianmei Wang; Yong Xu; Ali Winters; Linshu Wang; Xiaowei Dong; Eric Y Cheng; Ran Liu; Shao-Hua Yang
Journal:  PLoS One       Date:  2020-06-11       Impact factor: 3.240

3.  Quantitative analysis of seven plant hormones in Lotus japonicus using standard addition method.

Authors:  Takuyu Hashiguchi; Masatsugu Hashiguchi; Hidenori Tanaka; Koki Fukushima; Takahiro Gondo; Ryo Akashi
Journal:  PLoS One       Date:  2021-02-18       Impact factor: 3.240

4.  ExSTA: External Standard Addition Method for Accurate High-Throughput Quantitation in Targeted Proteomics Experiments.

Authors:  Yassene Mohammed; Jingxi Pan; Suping Zhang; Jun Han; Christoph H Borchers
Journal:  Proteomics Clin Appl       Date:  2017-10-25       Impact factor: 3.494

5.  Evaluation of Dissipation Behavior, Residues, and Dietary Risk Assessment of Fludioxonil in Cherry via QuEChERS Using HPLC-MS/MS Technique.

Authors:  Shunyu Yao; Zixi Zhao; Wang Lu; Xin Dong; Jiye Hu; Xiaolu Liu
Journal:  Molecules       Date:  2021-06-02       Impact factor: 4.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.