BACKGROUND AND OBJECTIVES: Conventional echocardiography is the technique of choice for assessing left ventricular function and the presence of structural heart disease in patients with heart failure. The aim of this study was to assess the diagnostic performance of handheld ultrasonography performed by a medical internist on patients with a clinical diagnosis of heart failure. PATIENTS AND METHODS: Cross-sectional observational study of 212 patients diagnosed with heart failure in a hospital center. A medical internist with basic training in echocardiography performed an examination using handheld ultrasonography and semiquantitatively assessed several variables. The patients' left ventricular systolic function was assessed, along with the cavity dimensions, significant valvular heart disease, pericardial effusion and the diameter of the inferior vena cava.
RESULTS: The examination using handheld ultrasonography was conducted in less than 6minutes. The agreement between the diagnoses of the medical internist and the expert echocardiographist was very good (k>0.81) for the diameter, hypertrophy and left ventricular systolic function, valvular regurgitation, pericardial effusion and diameter of the inferior vena cava.
CONCLUSIONS: Handheld echocardiography performed by a medical internist, as an extension of the physical examination of patients with heart failure, is a valid and safe test that helps increase the diagnostic performance.
Copyright © 2015 Elsevier España, S.L.U. y Sociedad Española de Medicina Interna (SEMI). All rights reserved.

Entities:  

Keywords:  Ecógrafo de bolsillo; Handheld ultrasound; Heart failure; Insuficiencia cardiaca; Left ventricle; Ventrículo izquierdo

Year:  2015        PMID: 25746465     DOI: 10.1016/j.rce.2015.01.015

Source DB:  PubMed          Journal:  Rev Clin Esp (Barc)        ISSN: 2254-8874


  1 in total

1.  Diagnostic accuracy of handheld cardiac ultrasound device for assessment of left ventricular structure and function: systematic review and meta-analysis.

Authors:  Sam Jenkins; Samer Alabed; Andrew Swift; Gabriel Marques; Alisdair Ryding; Chris Sawh; James Wardley; Benoy Nalin Shah; Peter Swoboda; Roxy Senior; Robin Nijveldt; Vassilios S Vassiliou; Pankaj Garg
Journal:  Heart       Date:  2021-08-06       Impact factor: 5.994

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.