| Literature DB >> 25743860 |
Sayoko Gotoh1, Daisuke Sugiyama1, Eriko Imai1, Mikito Kawamata1.
Abstract
A reinforced tracheal tube, ligated with silk threads, was inserted into a tracheostomy orifice and fixed to the skin. The cuff inflation line of the reinforced tracheal tube became occluded. Reinforced 'armoured' tracheal tubes have a spiral of wire embedded into the wall of the tube to give strength and flexibility, and may be sharply bent without compromising the tube lumen. The tracheal cuff attached to the tube is inflated by injecting air through a narrow-diameter tube welded to the outside of the tracheal tube. When a reinforced tracheal tube is ligated and fixed with silk threads, it should be confirmed whether the tracheal tube cuff can be deflated and inflated after fixation. Moreover, because occlusion can be eliminated by removing all silk threads used to ligate a tracheal tube, they should be removed before extubation. 2015 BMJ Publishing Group Ltd.Mesh:
Substances:
Year: 2015 PMID: 25743860 PMCID: PMC4369034 DOI: 10.1136/bcr-2014-207933