Literature DB >> 25740601

Force decay of latex and non-latex intermaxillary elastics: a clinical study.

Matheus Melo Pithon1, Jéssica Lima Mendes2, Camila Anselmo da Silva2, Rogério Lacerda Dos Santos2, Raildo da Silva Coqueiro3.   

Abstract

OBJECTIVE: To evaluate force decay between intermaxillary orthodontic elastics with and without latex in the oral medium at different time intervals.
MATERIALS AND METHODS: Intermaxillary orthodontic elastics with and without latex [1/8″ (3.2mm), 1/4″ (6.35mm), and 5/16″ (7.94mm) in diameter] were evaluated 'clinically'. For this purpose, 26 patients served as hosts for the elastics during the evaluation period. Initially, 52 thermoplasticized plates 1mm thick were made according to the patients' mould. Then, the orthodontic accessories were bonded to the plates, which kept the elastics distended throughout the experimental period. The force released by the elastics at this distance was evaluated at 0, 12, and 24 hours. Statistical analyses were performed using the Friedman, Wilcoxon, and Mann-Whitney tests (significance set at 5 per cent).
RESULTS: Elastics with and without latex presented with permanent deformation and increasing internal diameter throughout the entire period. Elastics with latex, 1/8″ (3.2mm) in diameter, sustained higher levels of force than those without latex. Elastics measuring 1/4″ (6.35mm) and 5/16″ (7.94mm) in diameter without latex sustained higher levels of force at 0 and 12 hours compared to those with latex. At the end of the experiment (24 hours), no significant difference was observed between the elastics (P > 0.05). LIMITATIONS: This study only evaluated three time intervals: 0, 12, and 24 hours. If the behaviour was evaluated at different time intervals, different behaviours of these materials might be revealed.
CONCLUSION: Elastics with latex 1/8″ in diameter sustained a higher level of final force than the corresponding type without latex. The 1/4 and 5/16 elastics with and without latex were not different at the end of the evaluation period.
© The Author 2015. Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25740601     DOI: 10.1093/ejo/cjv005

Source DB:  PubMed          Journal:  Eur J Orthod        ISSN: 0141-5387            Impact factor:   3.075


  3 in total

1.  The Effect of Extrinsic Factors on the Mechanical Behavior and Structure of Elastic Dental Ligatures and Chains.

Authors:  Kata Csekő; Péter Maróti; Zsuzsanna Helyes; Roland Told; Fanni Riegler; József Szalma; Zsuzsanna Gurdán
Journal:  Polymers (Basel)       Date:  2021-12-23       Impact factor: 4.329

2.  Comparison of the Force Released by Intermaxillary Elastics Used for Different Time Periods.

Authors:  Andressa Tribulato Lopes Nitrini; Adenilson Silva Chagas; Karina Maria Salvatore Freitas; Fabrício Pinelli Valarelli; Rodrigo Hermont Cançado; Renata Cristina Gobbi de Oliveira; Ricardo Cesar Gobbi de Oliveira
Journal:  Turk J Orthod       Date:  2019-12-01

3.  Sustainability in Orthodontics: what can we do to save our planet?

Authors:  Matheus Melo Pithon; Lucianne Cople Maia de Faria; Orlando Motohiro Tanaka; Antônio Carlos de Oliveira Ruellas; Laura Salignac de Souza Guimarães Primo
Journal:  Dental Press J Orthod       Date:  2017 Jul-Aug
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.