Effect of Aerosol Devices and Administration Techniques on Drug Delivery in a Simulated Spontaneously Breathing Pediatric Tracheostomy Model.

Bshayer R Alhamad1, James B Fink1, Robert J Harwood1, Meryl M Sheard1, Arzu Ari2.   

Abstract

BACKGROUND: This study was conducted to compare the efficiency of jet nebulizers, vibrating mesh nebulizers, and pressurized metered-dose inhalers (pMDI) during assisted and unassisted administration techniques using a simulated spontaneously breathing pediatric model with a tracheostomy tube (TT).
METHODS: An in vitro breathing model consisting of an uncuffed TT (4.5-mm inner diameter) was attached to a collecting filter (Respirgard) connected to a dual-chamber test lung and a ventilator (Hamilton Medical) to simulate breathing parameters of a 2-y-old child (breathing frequency, 25 breaths/min; tidal volume, 150 mL; inspiratory time, 0.8 s; peak inspiratory flow, 20 L/min). Albuterol sulfate was administered using a jet nebulizer (MicroMist, 2.5 mg/3 mL), vibrating mesh nebulizer (Aeroneb Solo, 2.5 mg/3 mL), and pMDI (ProAir HFA, 432 μg). Each device was tested 5 times with an unassisted technique (direct administration of aerosols with simulated spontaneous breathing) and with an assisted technique (using a manual resuscitation bag in conjunction with an aerosol device and synchronized with inspiration). Drug collected on the filter was analyzed by spectrophotometry.
RESULTS: With the unassisted technique, the pMDI had the highest inhaled mass percent (IM%, 47.15 ± 7.82%), followed by the vibrating mesh nebulizer (19.77 ± 2.99%) and the jet nebulizer (5.88 ± 0.77%, P = .002). IM was greater with the vibrating mesh nebulizer (0.49 ± .07 mg) than with the pMDI (0.20 ± 0.03 mg) and the jet nebulizer (0.15 ± 0.01 mg, P = .007). The trend of lower deposition with the assisted versus unassisted technique was not significant for the jet nebulizer (P = .46), vibrating mesh nebulizer (P = .19), and pMDI (P = .64).
CONCLUSIONS: In this in vitro pediatric breathing model with a TT, the pMDI delivered the highest IM%, whereas the vibrating mesh nebulizer delivered the highest IM. The jet nebulizer was the least efficient device. Delivery efficiency was similar with unassisted and assisted administration techniques.
Copyright © 2015 by Daedalus Enterprises.

Entities:  

Keywords:  aerosols; albuterol; drug delivery; metered dose inhaler; nebulizers; pediatrics; tracheostomy

Mesh:

Substances:

Year:  2015        PMID: 25737572     DOI: 10.4187/respcare.03592

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  In vitro evaluation of disposable transport ventilators with combination aerosol therapy.

Authors:  Marc Mac Giolla Eain; Andrew O'Sullivan; Mary Joyce; Ronan MacLoughlin
Journal:  BMJ Open Respir Res       Date:  2021-03

2.  In vitro delivery efficiencies of nebulizers for different breathing patterns.

Authors:  Hyun Mok Park; Kyung Hwa Chang; Sang-Hyub Moon; Bong Joo Park; Sun Kook Yoo; Ki Chang Nam
Journal:  Biomed Eng Online       Date:  2021-06-10       Impact factor: 2.819

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.