Ion mobility spectrometric analysis of vaporous chemical warfare agents by the instrument with corona discharge ionization ammonia dopant ambient temperature operation.

Takafumi Satoh1, Shintaro Kishi1, Hisayuki Nagashima1, Masumi Tachikawa2, Mieko Kanamori-Kataoka3, Takao Nakagawa4, Nobuyoshi Kitagawa4, Kenichi Tokita4, Soichiro Yamamoto4, Yasuo Seto5.   

Abstract

The ion mobility behavior of nineteen chemical warfare agents (7 nerve gases, 5 blister agents, 2 lachrymators, 2 blood agents, 3 choking agents) and related compounds including simulants (8 agents) and organic solvents (39) was comparably investigated by the ion mobility spectrometry instrument utilizing weak electric field linear drift tube with corona discharge ionization, ammonia doping, purified inner air drift flow circulation operated at ambient temperature and pressure. Three alkyl methylphosphonofluoridates, tabun, and four organophosphorus simulants gave the intense characteristic positive monomer-derived ion peaks and small dimer-derived ion peaks, and the later ion peaks were increased with the vapor concentrations. VX, RVX and tabun gave both characteristic positive monomer-derived ions and degradation product ions. Nitrogen mustards gave the intense characteristic positive ion peaks, and in addition distinctive negative ion peak appeared from HN3. Mustard gas, lewisite 1, o-chlorobenzylidenemalononitrile and 2-mercaptoethanol gave the characteristic negative ion peaks. Methylphosphonyl difluoride, 2-chloroacetophenone and 1,4-thioxane gave the characteristic ion peaks both in the positive and negative ion mode. 2-Chloroethylethylsulfide and allylisothiocyanate gave weak ion peaks. The marker ion peaks derived from two blood agents and three choking agents were very close to the reactant ion peak in negative ion mode and the respective reduced ion mobility was fluctuated. The reduced ion mobility of the CWA monomer-derived peaks were positively correlated with molecular masses among structurally similar agents such as G-type nerve gases and organophosphorus simulants; V-type nerve gases and nitrogen mustards. The slope values of the calibration plots of the peak heights of the characteristic marker ions versus the vapor concentrations are related to the detection sensitivity, and within chemical warfare agents examined the slope values for sarin, soman, tabun and nitrogen mustards were higher. Some CWA simulants and organic solvents gave the ion peaks eluting at the similar positions of the CWAs, resulting in false positive alarms.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Chemical warfare agents; Ion mobility spectrometry; Organic solvents; Performance evaluation; Simulants

Mesh:

Substances:

Year:  2015        PMID: 25732583     DOI: 10.1016/j.aca.2015.02.004

Source DB:  PubMed          Journal:  Anal Chim Acta        ISSN: 0003-2670            Impact factor:   6.558


  2 in total

1.  SAW Chemical Array Device Coated with Polymeric Sensing Materials for the Detection of Nerve Agents.

Authors:  Jinuk Kim; Hyewon Park; Jihyun Kim; Byung-Il Seo; Joo-Hyung Kim
Journal:  Sensors (Basel)       Date:  2020-12-08       Impact factor: 3.576

2.  A Potassium Ion-Exchanged Glass Optical Waveguide Sensor Locally Coated with a Crystal Violet-SiO2 Gel Film for Real-Time Detection of Organophosphorus Pesticides Simulant.

Authors:  Bin Du; Zhaoyang Tong; Xihui Mu; Jianjie Xu; Shuai Liu; Zhiwei Liu; Wei Cao; Zhi-Mei Qi
Journal:  Sensors (Basel)       Date:  2019-09-28       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.