[Purpose] The aim of this study was to investigate the clinical usefulness of a novel apparatus intended to improve the consistency and reliability of the deep tendon reflex (DTR) test. [Subjects] The participants included 50 healthy adults (25 male and 25 female) between 20 and 31 years of age who showed no knee abnormalities upon physical examination. [Methods] The intraclass correlation coefficient (ICC) was used to verify inter-rater reliability for two parameters examined in the DTR test using a newly designed apparatus. These parameters were the patellar reflex amplitude and first knee extension angle. Pearson's product correlation coefficient was then used to examine the correlation between these two parameters.Entities:
Keywords: Deep tendon reflex; First knee extension angle; Reliability
Year: 2015 PMID: 25729159 PMCID: PMC4339129 DOI: 10.1589/jpts.27.317
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Inter-tester reliability of the amplitude and FKEA of the deep tendon reflex test according to tapping angle (N=50)
| Tapping angle | 1st tester | 2nd tester | ICC(2,k) | 95% CI | ||
|---|---|---|---|---|---|---|
| Mean±SD | Range | Mean±SD | Range | |||
| Amplitude (mA) | ||||||
| 30° | 0.18±0.11 | 0.08–0.31 | 0.14±0.10 | 0.07–0.34 | 0.96 | 0.93–0.98 |
| 60° | 0.29±0.13 | 0.17–0.50 | 0.38±0.16 | 0.20–0.67 | 0.91 | 0.87–0.94 |
| 90° | 0.59±0.21 | 0.19–0.89 | 0.71±0.31 | 0.27–0.92 | 0.94 | 0.91–0.97 |
| 120° | 0.78±0.24 | 0.31–0.92 | 0.68±0.22 | 0.31–1.14 | 0.95 | 0.92–0.98 |
| FKEA (°) | ||||||
| 30° | 6.84±3.21 | 3.11–9.64 | 5.74±3.44 | 2.37–10.40 | 0.92 | 0.89–0.95 |
| 60° | 15.47±5.01 | 6.79–21.22 | 14.74±6.88 | 7.41–26.22 | 0.95 | 0.92–0.98 |
| 90° | 24.29±10.81 | 11.47–42.55 | 28.04±9.84 | 9.71–45.77 | 0.95 | 0.91–0.98 |
| 120° | 29.73±12.66 | 13.27–48.49 | 26.37±11.64 | 14.11–46.47 | 0.96 | 0.93–0.98 |
FKEA: first knee extension angle; SD: standard deviation; ICC: intraclass correlation coefficient; CI: confidence interval