Audit of radiation dose delivered in time-resolved four-dimensional computed tomography in a radiotherapy department.

Patricia Hubbard1, Jason Callahan2,3, Jim Cramb4, Ray Budd4, Tomas Kron3,4.   

Abstract

INTRODUCTION: To review the dose delivered to patients in time-resolved computed tomography (4D CT) used for radiotherapy treatment planning.
METHODS: 4D CT is used at Peter MacCallum Cancer Centre since July 2007 for radiotherapy treatment planning using a Philips Brilliance Wide Bore CT scanner (16 slice, helical 4D CT acquisition). All scans are performed at 140 kVp and reconstructed in 10 datasets for different phases of the breathing cycle. Dose records were analysed retrospectively for 387 patients who underwent 4D CT procedures between 2007 and 2013.
RESULTS: A total of 444 4D CT scans were acquired with the majority of them (342) being for lung cancer radiotherapy. Volume CT dose index (CTDIvol) as recorded over this period was fairly constant at approximately 20 mGy for adults. The CTDI for 4D CT for lung cancers of 19.6 ± 9.3 mGy (n = 168, mean ± 1SD) was found to be 63% higher than CTDIs for conventional CT scans for lung patients that were acquired in the same period (CTDIvol 12 ± 4 mGy, sample of n = 25). CTDI and dose length product (DLP) increased with increasing field of view; however, no significant difference between DLPs for different indications (breast, kidney, liver and lung) could be found. Breathing parameters such as breathing rate or pattern did not affect dose.
CONCLUSION: 4D CT scans can be acquired for radiotherapy treatment planning with a dose less than twice the one required for conventional CT scanning.
© 2015 The Royal Australian and New Zealand College of Radiologists.

Entities:  

Keywords:  imaging for treatment planning; motion management; radiation dose

Mesh:

Year:  2015        PMID: 25728211     DOI: 10.1111/1754-9485.12284

Source DB:  PubMed          Journal:  J Med Imaging Radiat Oncol        ISSN: 1754-9477            Impact factor:   1.735


  2 in total

1.  Independent review of 4DCT scans used for SABR treatment planning.

Authors:  Rachitha Antony; Peta Lonski; Elena Ungureanu; Nicholas Hardcastle; Adam Yeo; Shankar Siva; Tomas Kron
Journal:  J Appl Clin Med Phys       Date:  2020-02-13       Impact factor: 2.102

2.  Reducing radiation dose and enhancing imaging quality of 4DCT for radiation therapy using iterative reconstruction algorithms.

Authors:  George Noid; An Tai; Guang-Pei Chen; Jared Robbins; X Allen Li
Journal:  Adv Radiat Oncol       Date:  2017-04-22