Evaluating the effects of different dental devices on implantable cardioverter defibrillators.

Kanwal R Maheshwari1, Kian Nikdel1, Girard Guillaume2, Ariadne M Letra1, Renato M Silva1, Samuel O Dorn3.   

Abstract

INTRODUCTION: The implantable cardioverter defibrillator (ICD) is an electronic device that emits electrical signals to the heart via lead wires and electrodes. It is used for cardiac rhythm monitoring and treatment. Because electronic dental devices have been shown to produce electromagnetic fields, we hypothesize that they may interfere with ICD function.
METHODS: Nine dental devices (heat carrier, electronic apex locator, electric pulp tester, unipolar electrosurgery unit, electric motor, curing light, and 3 gutta-percha guns) were tested in this study for their ability to interfere with the function of 4 ICDs (2 single-chambered and 2 dual-chambered ICDs). ICD activity was monitored for 30 seconds using an ICD programmer (Medtronic 2090; Minneapolis, MN) and evaluated through an electrogram test strip printout.
RESULTS: Electromagnetic interference was detected with the electric motor, curing light, electric pulp tester, and electrosurgery unit although no electromagnetic disturbances were detected with these devices. No electromagnetic interferences were observed for the gutta-percha guns, heat carrier, and apex locator. However, the electrosurgery unit affected the dual-chambered ICD (Consulta CRT-D, Medtronic) and delivered therapies for fibrillation when no ventricular fibrillation was present.
CONCLUSIONS: Our results suggest that the electrosurgery unit produces electromagnetic disturbances with unwanted therapy delivery shock and potentially clinically significant outcomes.
Copyright © 2015 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Electromagnetic disturbance; electromagnetic interference; electronic dental devices; implantable cardioverter defibrillator

Mesh:

Year:  2015        PMID: 25726385     DOI: 10.1016/j.joen.2014.12.011

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  2 in total

Review 1.  Management of Patients With Cardiovascular Implantable Electronic Devices in Dental, Oral, and Maxillofacial Surgery.

Authors:  James Tom
Journal:  Anesth Prog       Date:  2016

2.  Influence and safety of electronic apex locators in patients with cardiovascular implantable electronic devices: a systematic review.

Authors:  Mothanna K AlRahabi; Hani M Ghabbani
Journal:  Libyan J Med       Date:  2019-12       Impact factor: 1.657

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.