| Literature DB >> 25723950 |
Rohanah Hussain1, Kristian Benning1, Tamas Javorfi1, Edoardo Longo1, Timothy R Rudd1, Bill Pulford1, Giuliano Siligardi1.
Abstract
The B23 Circular Dichroism beamline at Diamond Light Source has been operational since 2009 and has seen visits from more than 200 user groups, who have generated large amounts of data. Based on the experience of overseeing the users' progress at B23, four key areas requiring the most assistance are identified: planning of experiments and note-keeping; designing titration experiments; processing and analysis of the collected data; and production of experimental reports. To streamline these processes an integrated software package has been developed and made available for the users. The subsequent article summarizes the main features of the software.Entities:
Keywords: CDApps; denaturation; integrated software; protein; secondary structure; synchrotron radiation circular dichroism (SRCD); vacuum-ultraviolet (VUV)
Year: 2015 PMID: 25723950 PMCID: PMC4344363 DOI: 10.1107/S1600577514028161
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616
Figure 1Flow chart of the experimental protocol.
Figure 2Summary of CDApps presentations showing (A) the application page, (B) beam time visit information, (C) CD titration for ligand binding with K auto-fitting function, (D) CD measurement for thermal melt with auto-fitting by Boltzmann algorithm for determination, (E) protein secondary structure estimation using CONTINLL, SELCON3 or CDSSTR algorithms for batch processing, (F) project report generated in a Word document.
Beamline B23 CDApps features
| Features available | Notes |
|---|---|
| Link to the submitted project details | Requires the Oracle client to be installed |
|
| Based on molecular weight and concentration inputs from the user |
| Overlay of multiple spectra | Comparative viewing and assessment |
| Auto and manual fitting of the binding constant ( | Non-linear regression for single binding site |
| Stoichiometric determination | Indication of single or multiple binding sites |
| Auto and manual fitting of the melting temperature ( | Single Boltzmann algorithm |
| Secondary structure estimation for proteins and peptides | Auto fitting using various reference dataset libraries and three main algorithms: CONTINLL, CDSSTR or SELCON3 |
| Monitor the change in intensity at a single wavelength | Users can pick any wavelength value from a set of spectra to plot the changes in intensity as a function of time, temperature or scan number |
| Cope with data sets obtained with the six-cell turret | Data set can be split up to subsets belonging to the different samples in the turret and these subsets can be processed individually |
| Processed data and the analysis results are kept in various worksheets | Data are stored in an Excel file that can also be opened later on any computer |
| Report generation | Generates a pre-formatted Word document that includes the project number, name of users, abstract from peer-reviewed proposal, experimental details and selected figures; the discussion, conclusion, future work and references sections need to be filled in by the users |