Safety, reliability, and operability of cochlear implant electrode arrays coated with biocompatible polymer.

Makoto Kinoshita1, Yayoi S Kikkawa, Takashi Sakamoto, Kenji Kondo, Kazuhiko Ishihara, Tomohiro Konno, Nick Pawsey, Tatsuya Yamasoba.   

Abstract

CONCLUSION: Polymer-coated electrodes can reduce surgically-induced trauma associated with the insertion of a cochlear implant (CI) electrode array.
OBJECTIVES: To evaluate if insertion trauma in CI surgery can be reduced by using electrode arrays coated with 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer.
METHODS: We analyzed characteristics of the Contour Advance electrode arrays coated with MPC polymer. To assess surgical trauma during electrode insertion, polymer-coated or uncoated (n = 5 each) animal electrode arrays were implanted in guinea pig cochleae and operability and electrophysiological and histological changes were assessed.
RESULTS: Under light and scanning electron microscopy, polymer-coated electrodes did not appear different from uncoated electrodes, and no change was observed after mechanical stressing of the arrays. Electrode insertion was significantly easier when polymer-coated electrodes were used. Auditory brainstem response (ABR) thresholds did not differ between groups, but p1-n1 amplitudes of the coated group were larger compared with the uncoated group at 32 kHz at 28 days after surgery. The survival of outer hair cells and spiral ganglion cells was significantly greater in the polymer-coated group.

Entities:  

Keywords:  Auditory brainstem response; cochlear implantation; electrical acoustic stimulation; guinea pig; hair cell; organ of Corti; scanning electron microscope; spiral ganglion

Mesh:

Substances:

Year:  2015        PMID: 25719221     DOI: 10.3109/00016489.2014.990580

Source DB:  PubMed          Journal:  Acta Otolaryngol        ISSN: 0001-6489            Impact factor:   1.494


  3 in total

1.  Pharmacokinetics and Adverse Effects of 3 Sustained-release Buprenorphine Dosages in Healthy Guinea Pigs (Cavia porcellus).

Authors:  Andrea S Zanetti; Sumanth K Putta; Donald B Casebolt; Stan G Louie
Journal:  J Am Assoc Lab Anim Sci       Date:  2017-11-01       Impact factor: 1.232

2.  A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds.

Authors:  Dana Dohr; Katharina Wulf; Niels Grabow; Robert Mlynski; Sebastian P Schraven
Journal:  Materials (Basel)       Date:  2022-04-22       Impact factor: 3.748

3.  The impact of round window vs cochleostomy surgical approaches on interscalar excursions in the cochlea: Preliminary results from a flat-panel computed tomography study.

Authors:  Nicole T Jiam; Charles J Limb
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2016-11-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.