Seven Ventilators Challenged With Leaks During Neonatal Nasal CPAP: An Experimental Pilot Study.

Thomas Drevhammar1, Kjell Nilsson2, Henrik Zetterström3, Baldvin Jonsson4.   

Abstract

BACKGROUND: Nasal CPAP is the most common respiratory support for neonates. Several factors are considered important for effective treatment, including leaks at the patient interface and the delivery of pressure-stable CPAP. Investigations of pressure stability during leaks should include both the change in the mean delivered CPAP and the pressure variation during each breath. The aim of this study was to examine the response of ventilators delivering nasal CPAP when challenged with leaks at the patient interface.
METHODS: Seven ventilators providing nasal CPAP at 4 cm H2O were challenged with leaks during simulated neonatal breathing. Leak was applied for 15 consecutive breaths at a constant level (1-4 L/min).
RESULTS: The 2 aspects of pressure stability were evaluated by measuring the mean delivered CPAP and the amplitude of pressure swings before, during, and after leaks. The ability to maintain the delivered CPAP and the amplitude of pressure swings varied greatly among the 7 ventilators before, during, and after leaks. Four of the ventilators tested have built-in leak compensation.
CONCLUSIONS: There was no simple relationship between maintaining delivered CPAP during leaks and providing CPAP with low pressure swing amplitude. Maintaining the delivered CPAP and providing this without pressure swings are 2 separate aspects of pressure stability, and investigations concerning the clinical importance of pressure stability should address both aspects. This study also shows that compensation for leaks does not necessarily provide pressure-stable CPAP.
Copyright © 2015 by Daedalus Enterprises.

Entities:  

Keywords:  continuous positive airway pressure; equipment design; infant newborn; intensive care; mechanical; neonatal; ventilators; work of breathing

Mesh:

Year:  2015        PMID: 25715346     DOI: 10.4187/respcare.03718

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Electrical Activity of the Diaphragm in a Small Cohort of Preterm Infants on Noninvasive Neurally Adjusted Ventilatory Assist and Continuous Positive Airway Pressure: A Prospective Comparative Pilot Study.

Authors:  Arpit Gupta; Rishi Lumba; Sean Bailey; Sourabh Verma; Uday Patil; Pradeep Mally
Journal:  Cureus       Date:  2019-12-04

Review 2.  Aerosol drug delivery to spontaneously-breathing preterm neonates: lessons learned.

Authors:  Federico Bianco; Fabrizio Salomone; Ilaria Milesi; Xabier Murgia; Sauro Bonelli; Elena Pasini; Raffaele Dellacà; Maria Luisa Ventura; Jane Pillow
Journal:  Respir Res       Date:  2021-02-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.