Thomas Drevhammar1, Kjell Nilsson2, Henrik Zetterström3, Baldvin Jonsson4. 1. Department of Women's and Children's Health, Karolinska Institutet, Stockholm Department of Anaesthesiology, Östersund Hospital, Östersund thomas.drevhammar@regionjh.se. 2. Ventinvent AB, Östersund. 3. Department of Surgical Sciences, Anaesthesiology and Critical Care Medicine, Uppsala, Sweden. 4. Department of Women's and Children's Health, Karolinska Institutet, Stockholm.
Abstract
BACKGROUND: Nasal CPAP is the most common respiratory support for neonates. Several factors are considered important for effective treatment, including leaks at the patient interface and the delivery of pressure-stable CPAP. Investigations of pressure stability during leaks should include both the change in the mean delivered CPAP and the pressure variation during each breath. The aim of this study was to examine the response of ventilators delivering nasal CPAP when challenged with leaks at the patient interface. METHODS: Seven ventilators providing nasal CPAP at 4 cm H2O were challenged with leaks during simulated neonatal breathing. Leak was applied for 15 consecutive breaths at a constant level (1-4 L/min). RESULTS: The 2 aspects of pressure stability were evaluated by measuring the mean delivered CPAP and the amplitude of pressure swings before, during, and after leaks. The ability to maintain the delivered CPAP and the amplitude of pressure swings varied greatly among the 7 ventilators before, during, and after leaks. Four of the ventilators tested have built-in leak compensation. CONCLUSIONS: There was no simple relationship between maintaining delivered CPAP during leaks and providing CPAP with low pressure swing amplitude. Maintaining the delivered CPAP and providing this without pressure swings are 2 separate aspects of pressure stability, and investigations concerning the clinical importance of pressure stability should address both aspects. This study also shows that compensation for leaks does not necessarily provide pressure-stable CPAP.
BACKGROUND: Nasal CPAP is the most common respiratory support for neonates. Several factors are considered important for effective treatment, including leaks at the patient interface and the delivery of pressure-stable CPAP. Investigations of pressure stability during leaks should include both the change in the mean delivered CPAP and the pressure variation during each breath. The aim of this study was to examine the response of ventilators delivering nasal CPAP when challenged with leaks at the patient interface. METHODS: Seven ventilators providing nasal CPAP at 4 cm H2O were challenged with leaks during simulated neonatal breathing. Leak was applied for 15 consecutive breaths at a constant level (1-4 L/min). RESULTS: The 2 aspects of pressure stability were evaluated by measuring the mean delivered CPAP and the amplitude of pressure swings before, during, and after leaks. The ability to maintain the delivered CPAP and the amplitude of pressure swings varied greatly among the 7 ventilators before, during, and after leaks. Four of the ventilators tested have built-in leak compensation. CONCLUSIONS: There was no simple relationship between maintaining delivered CPAP during leaks and providing CPAP with low pressure swing amplitude. Maintaining the delivered CPAP and providing this without pressure swings are 2 separate aspects of pressure stability, and investigations concerning the clinical importance of pressure stability should address both aspects. This study also shows that compensation for leaks does not necessarily provide pressure-stable CPAP.
Authors: Federico Bianco; Fabrizio Salomone; Ilaria Milesi; Xabier Murgia; Sauro Bonelli; Elena Pasini; Raffaele Dellacà; Maria Luisa Ventura; Jane Pillow Journal: Respir Res Date: 2021-02-26