| Literature DB >> 25702273 |
Valentina Pirro1, Alan K Jarmusch2, Marco Vincenti3, R Graham Cooks4.
Abstract
Fourteen common drugs of abuse were identified in spiked oral fluid (ng mL(-1) levels), analyzed directly from medical swabs using touch spray mass spectrometry (TS-MS), exemplifying a rapid test for drug detection. Multiple stages of mass analysis (MS(2) and MS(3)) provided identification and detection limits sought by international forensic and toxicological societies, Δ(9)-THC and buprenorphine excluded. The measurements were made using a medical swab as both the sampling probe and means of ionization. The adaptation of medical swabs for TS-MS analysis allows non-invasive and direct sampling of neat oral fluid. Data acquisition was rapid, seconds per drug, and MS(3) ensured reliable identification of illicit drugs. The reported data were acquired to investigate (i) ionization of common drugs from commercial swabs, (ii) ion intensity over spray duration, and (iii) dynamic range, all as initial steps in development of a quantitative method. The approach outlined is intended for point-of-care drug testing using oral fluid in clinical applications as well as in situ settings, viz. in forensic applications. The proof-of-concept results presented will require extension to other controlled substances and refinement in analytical procedures to meet clinical/legal requirements.Entities:
Keywords: Ambient ionization; Drug testing; Medical swab; Tandem mass spectrometry; Touch spray ionization; Toxicology
Mesh:
Substances:
Year: 2015 PMID: 25702273 PMCID: PMC4513665 DOI: 10.1016/j.aca.2015.01.008
Source DB: PubMed Journal: Anal Chim Acta ISSN: 0003-2670 Impact factor: 6.558