Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions.

A Asena1, T Kairn2, S B Crowe3, J V Trapp3.   

Abstract

PURPOSE: This study investigates the effects of temporary tissue expanders (TTEs) on the dose distributions in breast cancer radiotherapy treatments under a variety of conditions.
METHODS: Using EBT2 radiochromic film, both electron and photon beam dose distribution measurements were made for different phantoms, and beam geometries. This was done to establish a more comprehensive understanding of the implant's perturbation effects under a wider variety of conditions.
RESULTS: The magnetic disk present in a tissue expander causes a dose reduction of approximately 20% in a photon tangent treatment and 56% in electron boost fields immediately downstream of the implant. The effects of the silicon elastomer are also much more apparent in an electron beam than a photon beam.
CONCLUSIONS: Evidently, each component of the TTE attenuates the radiation beam to different degrees. This study has demonstrated that the accuracy of photon and electron treatments of post-mastectomy patients is influenced by the presence of a tissue expander for various beam orientations. The impact of TTEs on dose distributions establishes the importance of an accurately modelled high-density implant in the treatment planning system for post-mastectomy patients.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Dose distribution; Film; High-density; Magnet; Tissue expander

Mesh:

Year:  2015        PMID: 25693908     DOI: 10.1016/j.ejmp.2015.01.015

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Dosimetric analysis of the effects of a temporary tissue expander on the radiotherapy technique.

Authors:  So Hyun Park; Young Suk Kim; Jinhyun Choi
Journal:  Radiol Med       Date:  2020-10-06       Impact factor: 3.469

2.  Robustness of three external beam treatment techniques against inter-fractional positional variations of the metal port in breast tissue expanders.

Authors:  Keren Mayorov; Patricia Lacasse; Elsayed Ali
Journal:  J Appl Clin Med Phys       Date:  2021-11-22       Impact factor: 2.102

3.  Critical Factors of Dose Distribution in Breast Cancer Tomotherapy With Metallic Port Breast Tissue Expander: Image Correction, Delivery Mode, and Volume Impact.

Authors:  Hsing-Yi Lee; Yu-Hsiu Yen; Yu-Lun Tsai; Pei-Chih Tu; Chi-Ming Pu; Chia-Hong Lin; Louis Tak Lui; Suzun Shaw; Ching-Jung Wu; Hsin-Hua Nien
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.