Pre-clinical validation study of a miniaturized electrochemical immunoassay based on square wave voltammetry for early detection of carcinoembryonic antigen in human serum.

Flavio Dolores Martínez-Mancera1, Patricia García-López2, José Luis Hernández-López3.   

Abstract

BACKGROUND: The ELISA format for measuring carcinoembryonic antigen (CEA) serves as a reference standard against which other assays are compared. Because the World Health Organization (WHO) increasingly recommends the use of serum CEA as a diagnostic tool for cancer, it is relevant to explore the reliability of the new decentralized CEA point-of-care-testing (POCT) technologies that are available to physicians and patients, in compliance with mandates of the clinical laboratories' regulatory agencies.
METHODS: Electrochemical immunoassay (ECIA) based on trace lead (Pb) analysis by anodic stripping techniques using sandwich-type immunocomplex conjugates: (MB)Ab/AgCEA/Ab(PbS), and a commercial ELISA test system with optical transmission.
RESULTS: The ECIA provides better analytical performance than does the ELISA. The within assay precision coefficient of variance (%CVw) of the ECIA is lower than the value recommended by the Hong Kong Association of Medical Laboratories (HKAML), and the recoveries of CEA at 1.0, 5.0, 10.0, 25.0 and 50.0 ng/ml are in the range of 99-110% for control serum samples. The ECIA showed a minimal positive bias of 0.0267 ± 0.3270 ng/ml (P=0.9389).
CONCLUSIONS: The proposed CEA screening technology can be practically employed for decentralized clinical analysis of CEA in human serum. Therefore, it can be viewed as a control method for personalized therapy.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Biostatistical analysis; Carcinoembryonic antigen; ELISA test system; Magnetic beads; Miniaturization; Square wave voltammetry

Mesh:

Substances:

Year:  2015        PMID: 25689793     DOI: 10.1016/j.cca.2015.02.017

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  5 in total

1.  Portable electrochemical immunosensor for detection of Mycobacterium tuberculosis secreted protein CFP10-ESAT6 in clinical sputum samples.

Authors:  Umi Zulaikha Mohd Azmi; Nor Azah Yusof; Jaafar Abdullah; Shahrul Ainliah Alang Ahmad; Fatin Nabilah Mohd Faudzi; Nurul Hanun Ahmad Raston; Siti Suraiya; Poh Shing Ong; Devandran Krishnan; Nur Khairunnisa Sahar
Journal:  Mikrochim Acta       Date:  2021-01-06       Impact factor: 5.833

2.  Exonuclease III-assisted fluorometric aptasensor for the carcinoembryonic antigen using graphene oxide and 2-aminopurine.

Authors:  Mingjian Chen; Changbei Ma; Han Zhao; Ying Yan
Journal:  Mikrochim Acta       Date:  2019-07-03       Impact factor: 5.833

3.  A Nuclease Protection ELISA Assay for Colorimetric and Electrochemical Detection of Nucleic Acids.

Authors:  Jessica E Filer; Robert B Channon; Charles S Henry; Brian J Geiss
Journal:  Anal Methods       Date:  2019-01-24       Impact factor: 2.896

4.  Electrochemical sensor for rapid determination of fibroblast growth factor receptor 4 in raw cancer cell lysates.

Authors:  Rebeca M Torrente-Rodríguez; Víctor Ruiz-Valdepeñas Montiel; Susana Campuzano; María Pedrero; Meryem Farchado; Eva Vargas; F Javier Manuel de Villena; María Garranzo-Asensio; Rodrigo Barderas; José M Pingarrón
Journal:  PLoS One       Date:  2017-04-04       Impact factor: 3.240

Review 5.  Recent Developments of Electrochemical and Optical Biosensors for Antibody Detection.

Authors:  Wei Xu; Daniel Wang; Derek Li; Chung Chiun Liu
Journal:  Int J Mol Sci       Date:  2019-12-24       Impact factor: 5.923

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.