Literature DB >> 25686908

Electron microscopic evaluation of a gold glaucoma micro shunt after explantation.

Thomas A Berk1, Diamond Y Tam1, Liliana Werner1, Nick Mamalis1, Iqbal Ike K Ahmed2.   

Abstract

UNLABELLED: We present a case of an explanted gold glaucoma micro shunt (GMS Plus) and the subsequent light and electron microscopic analyses. The shunt was implanted in a patient with medically refractive glaucoma. The intraocular pressure (IOP) was stable at 12 mm Hg 6 months postoperatively but spiked to 26 mm Hg 6 months later; membranous growth was visible on the implant gonioscopically. A second gold micro shunt was placed 2 years after the first. The IOP was 7 mm Hg 1 week postoperatively but increased to 23 mm Hg 3 weeks later; similar membranous growth was visible on this implant. One of the shunts was explanted, and light and scanning electron microscopic analyses revealed encapsulation around the shunt exterior and connective tissue invasion of the microstructure. This represents the first electron microscopic analysis of an explanted gold glaucoma micro shunt and the first unequivocal images of the fibrotic pseudo-capsule traversing its microchannels and fenestrations. FINANCIAL DISCLOSURE: Dr. Ahmed is a consultant to and has received research grants from Solx, Inc. No other author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25686908     DOI: 10.1016/j.jcrs.2014.12.001

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Safety and effectiveness of gold glaucoma micro shunt for reducing intraocular pressure in Japanese patients with open angle glaucoma.

Authors:  Masaki Tanito; Etsuo Chihara
Journal:  Jpn J Ophthalmol       Date:  2017-06-09       Impact factor: 2.447

Review 2.  The Suprachoroidal Route in Glaucoma Surgery.

Authors:  Anthony Gigon; Tarek Shaarawy
Journal:  J Curr Glaucoma Pract       Date:  2016-05-12

3.  Gold shunt for refractory advanced low-tension glaucoma with spared central acuity.

Authors:  Ryan Le; Neeru Gupta
Journal:  Int Med Case Rep J       Date:  2016-03-18

Review 4.  Translating Minimally Invasive Glaucoma Surgery Devices.

Authors:  Richard M H Lee; Yann Bouremel; Ian Eames; Steve Brocchini; Peng Tee Khaw
Journal:  Clin Transl Sci       Date:  2019-09-30       Impact factor: 4.689

5.  Ultrastructural analysis of explanted CyPass microstents and correlation with clinical findings.

Authors:  Lisa Hübner; U Schlötzer-Schrehardt; J M Weller; B Hohberger; C Y Mardin; R Lämmer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-03-09       Impact factor: 3.535

6.  Long-term Follow-up of a Case of Gold Shunt Surgery for Refractory Silicone Oil-induced Glaucoma.

Authors:  Ryan Le; Alan R Berger; Neeru Gupta
Journal:  J Glaucoma       Date:  2016-08       Impact factor: 2.503

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.