Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3.

Davide Cusumano1, Maria L Fumagalli2, Marcello Marchetti3, Laura Fariselli3, Elena De Martin2.   

Abstract

Aim of this study is to examine the feasibility of using the new Gafchromic EBT3 film in a high-dose stereotactic radiosurgery and radiotherapy quality assurance procedure. Owing to the reduced dimensions of the involved lesions, the feasibility of scanning plan verification films on the scanner plate area with the best uniformity rather than using a correction mask was evaluated. For this purpose, signal values dispersion and reproducibility of film scans were investigated. Uniformity was then quantified in the selected area and was found to be within 1.5% for doses up to 8 Gy. A high-dose threshold level for analyses using this procedure was established evaluating the sensitivity of the irradiated films. Sensitivity was found to be of the order of centiGray for doses up to 6.2 Gy and decreasing for higher doses. The obtained results were used to implement a procedure comparing dose distributions delivered with a CyberKnife system to planned ones. The procedure was validated through single beam irradiation on a Gafchromic film. The agreement between dose distributions was then evaluated for 13 patients (brain lesions, 5 Gy/die prescription isodose ~80%) using gamma analysis. Results obtained using Gamma test criteria of 5%/1 mm show a pass rate of 94.3%. Gamma frequency parameters calculation for EBT3 films showed to strongly depend on subtraction of unexposed film pixel values from irradiated ones. In the framework of the described dosimetric procedure, EBT3 films proved to be effective in the verification of high doses delivered to lesions with complex shapes and adjacent to organs at risk.
Copyright © 2015 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dosimetric verification; Gafchromic EBT3; High dose; Stereotactic radiosurgery/radiotherapy

Mesh:

Year:  2015        PMID: 25680475     DOI: 10.1016/j.meddos.2015.01.001

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  4 in total

1.  Dosimetric comparison of different treatment modalities for stereotactic radiotherapy.

Authors:  Shih-Ming Hsu; Yuan-Chun Lai; Chien-Chung Jeng; Chia-Ying Tseng
Journal:  Radiat Oncol       Date:  2017-09-16       Impact factor: 3.481

2.  Sum signal dosimetry: A new approach for high dose quality assurance with Gafchromic EBT3.

Authors:  Davide Cusumano; Maria Luisa Fumagalli; Francesco Ghielmetti; Linda Rossi; Giuliano Grossi; Raffaella Lanzarotti; Laura Fariselli; Elena De Martin
Journal:  J Appl Clin Med Phys       Date:  2017-02-02       Impact factor: 2.102

3.  Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array.

Authors:  Juan Fernando Mata Colodro; Alfredo Serna Berná; Vicente Puchades Puchades; David Ramos Amores; Miguel Alcaraz Baños
Journal:  J Med Phys       Date:  2017 Jul-Sep

4.  A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Sujatha Manikandan
Journal:  J Med Phys       Date:  2018 Apr-Jun
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.