| Literature DB >> 2566674 |
V S Chungi1, G S Rekhi, L Shargel.
Abstract
A simple and rapid assay for quantitation of sulfasalazine metabolites in rat urine and plasma was developed using high-performance liquid chromatography (HPLC). The method involves dilution of urine or plasma samples (0.1 mL) with methanol for protein precipitation, followed by mixing and centrifugation at 10,000 x g. Chromatography was accomplished with a reversed-phase ODS C-18 column (5 mu; 4.6 x 250 mm). The mobile phase consisted of 20% methanol in 5.0 mM phosphate buffer (pH 6.0), with 0.5 mM tetrabutylammonium chloride as an ion-pairing agent. The flow rate was 1.7 mL/min. An injection volume of 30 microL was used and the metabolites were quantitated by an ultraviolet detector at 254 nm. Benzamide was used as the internal standard. This method is linear in the range of 0.5 to 25 micrograms/mL for 5-aminosalicylic acid (5-ASA), acetylsulfapyridine (Ac-SP), and acetyl-5-aminosalicylic acid (Ac-5-ASA), and from 0.25 to 25 micrograms/mL for sulfapyridine (SP). The percent relative standard deviation ranged from 1 to 7.9% for the metabolite standard curves and precision studies. The limit of detection for 5-ASA, Ac-SP, and Ac-5-ASA is 100 ng/mL, and for SP is 50 ng/mL, in both urine and plasma. This method is rapid, precise, and accurate, and has been used to determine sulfasalazine metabolites in individual rat plasma and urine samples following an oral dose of 60 mg/kg of sulfasalazine.Entities:
Mesh:
Substances:
Year: 1989 PMID: 2566674 DOI: 10.1002/jps.2600780313
Source DB: PubMed Journal: J Pharm Sci ISSN: 0022-3549 Impact factor: 3.534