[The biological safety evaluation of a new developed silicone rubber for inflatable silastic prosthesis].

Tian Ai, Liang Xing, Liao Jian, Sun Xu, Teng Minhua, Zhang Li, Chen Yue.   

Abstract

OBJECTIVE: The biological safety of a new developed silicone rubber for inflatable silastic prosthesis (SRISP) was evaluated.
METHODS: Following the GB/T 16886.10-2005 standard, YY/T 0127.13-2009 standard, and GB/T 16886.11- 2011 standard, samples were prepared and tested by animal experiments, such as guinea pig maximization test, oral mucous membrane irritation test, and short-term systemic toxicity test (oral route).
RESULTS: No obvious erythema and edema in the guinea pig abdominal skin were observed after 24, 48, and 72 h of stimulating touch, thus indicating that SRISP does not cause potential skin sensitivity. No local response to SRISP was found, and the visual observation and pathological findings of oral mucosa were normal and similar to that of the control group. Therefore, SRISP had no irritation response to oral mucosa. No clinical signs of toxicity were observed in rats, and no significant differences in weight and weight relative growth rate between extract group and blank control group (P > 0.05) were found. Thus, SRISP had no short-term systemic toxicity.
CONCLUSION: These results indicated that SRISP met the requirement of biomedical materials and had good bio- security.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25665419      PMCID: PMC7030702     

Source DB:  PubMed          Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi        ISSN: 1000-1182


  6 in total

Review 1.  [Progress in the studies on the evaluation of biocompatibility of biomaterials].

Authors:  X Yang; T Xi
Journal:  Sheng Wu Yi Xue Gong Cheng Xue Za Zhi       Date:  2001-03

2.  An elastomeric material for facial prostheses: synthesis, experimental and numerical testing aspects.

Authors:  Kathryn Bellamy; G Limbert; Mark G Waters; J Middleton
Journal:  Biomaterials       Date:  2003-12       Impact factor: 12.479

3.  Health-related quality of life after maxillectomy: a comparison between prosthetic obturation and free flap.

Authors:  Simon N Rogers; Derek Lowe; Deborah McNally; James S Brown; E David Vaughan
Journal:  J Oral Maxillofac Surg       Date:  2003-02       Impact factor: 1.895

Review 4.  Fluoride-containing gels, mouth rinses and varnishes: an update of evidence of efficacy.

Authors:  S Poulsen
Journal:  Eur Arch Paediatr Dent       Date:  2009-09

Review 5.  Patient satisfaction with maxillofacial prosthesis. Literature review.

Authors:  Marcelo Coelho Goiato; Aldiéris Alves Pesqueira; Cristina Ramos da Silva; Humberto Gennari Filho; Daniela Micheline Dos Santos
Journal:  J Plast Reconstr Aesthet Surg       Date:  2008-12-06       Impact factor: 2.740

6.  Effect of multi-adhesive layering on retention of extraoral maxillofacial silicone prostheses in vivo.

Authors:  Sudarat Kiat-Amnuay; Lawrence Gettleman; L Jane Goldsmith
Journal:  J Prosthet Dent       Date:  2004-09       Impact factor: 3.426

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.