| Literature DB >> 25662018 |
Nicolas Salez1, Antoine Nougairede2, Laetitia Ninove2, Christine Zandotti2, Xavier de Lamballerie2, Remi N Charrel2.
Abstract
A total of 281 clinical specimens (nasal swabs and nasopharyngeal aspirates) were tested with the Xpert® Flu/RSV XC. The results were compared to those obtained with the real-time retro transcriptase-polymerase chain reaction assays routinely used in our laboratory. The Xpert® Flu/RSV XC showed sensitivity/specificity of 97.8%/100% and 97.9%/100% for flu and respiratory syncytial virus, respectively.Entities:
Keywords: Diagnosis; Flu; Point-of-care; RT-PCR; Rapid Influenza Detection Test; Real-time PCR; Syncytial respiratory virus
Mesh:
Year: 2015 PMID: 25662018 PMCID: PMC7172894 DOI: 10.1016/j.diagmicrobio.2015.01.008
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Comparative evaluation of Cepheid Xpert® Flu/RSV XC assay and routine rt-RT-PCR tests on 281 selected specimens.
| Type of storage | Result category | Flu A H3N2 (n = 39) | Flu A H1N1 (n = 28) | Flu B (n = 22) | RSV A (n = 55) | RSV B (n = 42) | hMPV (n = 20) | hRV (n = 54) | hCoV (n = 14) | Negative (N = 7) |
|---|---|---|---|---|---|---|---|---|---|---|
| Fresh unfrozen samples prospectively tested | Positive with the Real time RT-PCR assay (reference test) | 18 | 4 | - | 2 | 2 | 6 | 10 | 1 | 2 |
| Positive with the Xpert® Flu/RSV assay | 18 | 4 | - | 2 | 1 | 0 | 0 | 0 | 0 | |
| Frozen samples retrospectively tested | Positive with the rt-RT-PCR assay (reference test) | 21 | 24 | 22 | 53 | 40 | 14 | 44 | 13 | 5 |
| Positive with the Xpert® Flu/RSV assay | 20 | 23 | 22 | 53 | 39 | 0 | 0 | 0 | 0 |