| Literature DB >> 25650079 |
M P Quílez1, M A Pérez2, B Seral-García3.
Abstract
The best management of severe bone defects following total knee replacement is still controversial. Metal augments, tantalum cones and porous tibial sleeves could help the surgeon to manage any type of bone loss, providing a stable and durable knee joint reconstruction. Five different types of prostheses have been analysed: one prosthesis with straight stem; two prostheses with offset stem, with and without supplement, and two prostheses with sleeves, with and without stem. The purpose of this study is to report a finite element study of revision knee tibial implants. The main objective was to analyse the tibial bone density changes and Von Misses tension changes following different tibial implant designs. In all cases, the bone density decreases in the proximal epiphysis and medullary channels, with a bone density increase also being predicted in the diaphysis and at the bone around the stems tips. The highest value of Von Misses stress has been obtained for the straight tibial stem, and the lowest for the stemless metaphyseal sleeves prosthesis.Entities:
Keywords: Bone resorption; Elementos finitos; Finite element; Metaphyseal sleeves; Reabsorción ósea; Recambio de prótesis de rodilla; Revision total knee arthroplasty; Stems; Vainas metafisarias; Vástagos
Mesh:
Year: 2015 PMID: 25650079 DOI: 10.1016/j.recot.2014.12.003
Source DB: PubMed Journal: Rev Esp Cir Ortop Traumatol ISSN: 1888-4415