Efficiency of short, small-diameter columns for reversed-phase liquid chromatography under practical operating conditions.

Yan Ma1, Alexander W Chassy1, Shota Miyazaki2, Masanori Motokawa2, Kei Morisato3, Hideyuki Uzu2, Masayoshi Ohira2, Masahiro Furuno2, Kazuki Nakanishi4, Hiroyoshi Minakuchi5, Khaled Mriziq6, Tivadar Farkas7, Oliver Fiehn8, Nobuo Tanaka9.   

Abstract

Prototype small-size (1.0mm I.D., 5cm long) columns for reversed-phase HPLC were evaluated in relation to instrument requirements. The performance of three types of columns, monolithic silica and particulate silica (2μm, totally porous and 2.6μm, core-shell particles) was studied in the presence of considerable or minimal extra-column effects, while the detector contribution to band broadening was minimized by employing a small size UV-detector cell (6- or 90nL). A micro-LC instrument having small system volume (<1μL) provided extra-column band variance of only 0.01-0.02μL(2). The three columns generated about 8500 theoretical plates for solutes with retention factor, k>1-3 (depending on the column), in acetonitrile/water mobile phase (65/35=vol/vol) at 0.05mL/min, with the instrument specified above. The column efficiency was lower by up to 30% than that observed with a 2.1mm I.D. commercial column. The small-size columns also provided 8000-8500 theoretical plates for well retained solutes with a commercial ultrahigh-pressure liquid chromatography (UHPLC) instrument when extra-column contributions were minimized. While a significant extra-column effect was observed for early eluting solutes (k<2-4, depending on column) with methanol/water (20/80=vol/vol) as weak-wash solvent, the use of methanol/water=50/50 as wash solvent affected the column efficiency for most analytes. The results suggest that the band compression effect by the weak-wash solvent associated with partial-loop injection may provide a practical means to reducing the extra-column effect for small-size columns, while the use of an instrument with minimum extra-column effect is highly desirable.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  1mm I.D. column; Column efficiency; Core–shell particles; Extra-column effect; Monolithic silica; Weak-wash solvent

Mesh:

Substances:

Year:  2015        PMID: 25648581     DOI: 10.1016/j.chroma.2015.01.014

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Ultrafast Polyphenol Metabolomics of Red Wines Using MicroLC-MS/MS.

Authors:  Yan Ma; Nobuo Tanaka; Arpana Vaniya; Tobias Kind; Oliver Fiehn
Journal:  J Agric Food Chem       Date:  2016-01-05       Impact factor: 5.279

2.  Development of Narrow-Bore C18 Column for Fast Separation of Peptides and Proteins in High-Performance Liquid Chromatography.

Authors:  Ashraf Ali; Sarah Alharthi; Nora Hamad Al-Shaalan; Eman Y Santali
Journal:  Polymers (Basel)       Date:  2022-06-25       Impact factor: 4.967

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.