Etsuko Amakishi1, Yoriko Irie, Kanae Nishizawa, Hiromi Kamada, Fumiaki Nakajima, Nobuki Matsuyama, Hiroyuki Ishii, Harumichi Matsukura, Kazuta Yasui, Fumiya Hirayama.
Abstract
Granulocyte immunofluorescence and granulocyte agglutination tests are standard methods for detecting human neutrophil antigen (HNA) antibodies (Abs); however, these require a typed panel of neutrophils, which can be time-consuming to develop, and it remains difficult to determine antibody specificity in some cases. We established and evaluated four detection systems for HNA-1a Abs based on an HNA-1a-expressing cell line (KY cells) and antigen capture. We additionally evaluated a commercial solid-phase system. Eleven HNA-1a antibody-positive samples, including the World Health Organization Reference Reagent, and 40 serum samples derived from male blood donors were used as positive and negative control samples, respectively. Although specificity was >0.90 in all systems evaluated, the sensitivity varied among the systems. The KY cell-based monoclonal antibody specific immobilisation of granulocyte antigens (KY-MAIGA) system using certain, but not all, monoclonal Abs, and the solid-phase system revealed higher sensitivity than other systems. In conclusion, the KY-MAIGA and commercial solid-phase systems were superior in terms of specific and sensitive detection of HNA-1a Abs.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25644150 DOI: 10.1007/s12185-015-1751-y
Source DB: PubMed Journal: Int J Hematol ISSN: 0925-5710 Impact factor: 2.490