Literature DB >> 25639130

Assessment of a portable device to quantify vitamin A in fortified foods (flour, sugar, and milk) for quality control.

Arnaud Laillou, Cécile Renaud, Jacques Berger, Regina Moench-Pfanner, Laura Fontan, Sylvie Avallone.   

Abstract

BACKGROUND: Simple-to-use quantitative methods are needed to check the adequacy of vitamin A fortification levels.
OBJECTIVE: To assess the capacity of a portable fluorometer (iCheck FLUORO) and its test kit vials (iEx Mila) to quantify retinyl palmitate in fortified milks, flours (wheat, maize), and sugar.
METHODS: The portable fluorometer was assessed in a three-step procedure to determine its working range and linearity, intra-assay precision, and interperson precision. Measurements were compared with the results obtained by high-performance liquid chromatography (HPLC), commonly regarded as the standard method for vitamin A analysis.
RESULTS: The portable fluorometer (iCheck FLUORO) and its test kit vials (iEx Mila) precisely determined the vitamin A contents in fortifed flours, sugar, and milks. Its working range was 1 to 10, 0.5 to 3.0, and 5 to 15 mg retinol equivalents (RE) kg(-1) for flours (wheat and corn), milks, and sugar, respectively; these values are in accordance with the World Health Organization recommendations for food fortification in least developed countries. The limits of detection are higher than those of HPLC but are all satisfactory (< 1.46 mg RE kg(-1)). The coefficients of variation within and between observers were satisfactory, especially for sugar and milk.
CONCLUSIONS: The linear relationship between the data from the portable fluorometer and the HPLC data confirms that the portable fluorometer provides a good determination of the vitamin A content of the fortified products in the tested range.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25639130     DOI: 10.1177/156482651403500407

Source DB:  PubMed          Journal:  Food Nutr Bull        ISSN: 0379-5721            Impact factor:   2.069


  2 in total

1.  Validation of blood vitamin A concentrations in cattle: comparison of a new cow-side test (iCheck™ FLUORO) with high-performance liquid chromatography (HPLC).

Authors:  Jens Raila; Chiho Kawashima; Helga Sauerwein; Nadine Hülsmann; Christoph Knorr; Akio Myamoto; Florian J Schweigert
Journal:  BMC Vet Res       Date:  2017-05-10       Impact factor: 2.741

2.  A comparison study of five different methods to measure carotenoids in biofortified yellow cassava (Manihot esculenta).

Authors:  Angélica M Jaramillo; Luis Fernando Londoño; Juan Camilo Orozco; Gelver Patiño; John Belalcazar; Fabrice Davrieux; Elise F Talsma
Journal:  PLoS One       Date:  2018-12-28       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.